3
defining the valve alignment position and the working length of the balloon. A radiopaque double marker
proximal to the balloon indicates flex catheter position during deployment. The inflation parameters for
valve deployment are:
Table 4
Model
Nominal Balloon
Diameter
Nominal Inflation
Volume
Rated Burst
Pressure (RBP)
9355FS23
23 mm
17 mL
7 atm
9355FS26
26 mm
22 mL
7 atm
9355FS29
29 mm
33 mL
7 atm
Figure 2a. N Delivery System
Figure 2b. Default Position
Figure 2c. Valve Alignment Position
•
Qualcrimp Crimping Accessory (Figure 3)
The Qualcrimp crimping accessory (packaged with the N delivery system and manufactured in
laminated or cloth covered foam) is used during crimping of the valve.
Figure 3
Laminated Qualcrimp
Cloth Qualcrimp