will open when the normal signal becomes active (pump at speed)
▪
will close when you select the stop button or if there is a fault condition
The reaction time will be in line with the valve selection and the output signal is 24 V d.c.
Refer to
for polarity of the connector pins when the connector is
energized.
The auxiliary connector is regulated to 24 V d.c. to control the accessories. If the
auxiliary load current exceeds the value in
Table: Auxiliary load currents
, the output will
shut down to protect the pump controller. Refer to
Table: Recommended mating plugs
for the recommended mating plugs.
Figure 3 Valve connector
Pin number
Signal
Polarity
1
Valve 1 - Control output
Positive
2
Valve 2 - Control output
Positive
3
Valve 1 - Return
Negative
4
Valve 2 - Return
Negative
Table 1
Auxiliary load currents
Description
Data
Connector plug
Phoenix part number SACC-DSI-M 8FS-4CON-M12/0.5
Voltage output
24 V d.c. -25%, +10% (18 V d.c. to 26.4 V d.c.)
Output power
2 Channels with 4 watts per channel
Table 2
Recommended mating cable
Mating connector cable
Phoenix part number
Screw connection, straight
SACC-M 8MS-4CON-M-SW
Solder connection, straight
SACC-M 8MS-4CON-M
Screw connection, right angle
SACC-M 8MR-4CON-M-SW
Solder connection, right angle
SACC-M 8MR-4CON-M
2.4 Logic interface
The logic interface is designed to support the serial control, the parallel control, monitor
and operate through one connector.