Object name Command Parameter
range
Factory
setting Data type
Units
Comments
Pump cycles
?V884
0..65535
-
decimal
cycles
Cycles run by pump
0..65535
decimal
cycles
Cycles until pump service due
Bearing run
time
?V885
0..999999
-
decimal
hours
Hours run by bearing
0..999999
decimal
hours
Hours until bearing service due
Oil cartridge
run time
?V886
0..999999
-
decimal
hours
Hours run by oil cartridge
0..999999
decimal
hours
Hours until oil cartridge service
due
Table 14
Command abbreviations
Abbreviation
Meaning
cr
carriage return character
chars
characters
d
decimal ASCII character
h
hexadecimal ASCII character
r
Returned error code. Refer to
commands that can be sent to the nEXT pump on page
sp
space character
string
can have several ASCII characters
X
multi-drop decimal ASCII character
Note:
Fields that have multiple 'd' characters shows the typical length. All data fields have a
maximum of 5 decimal characters (prefixed by a minus number for negative numbers).
Fields that have multiple 'X' characters shows the maximum length and does not show
the fixed length.
Table 15
Error codes
Returned error code
Meaning
0
No error
1
Invalid command for object ID
2
Invalid Query or Command
3
Missing parameter
4
Parameter out of range
5
Invalid command in current state - Example: serial
command to start or stop when in parallel control
mode
Page 49
B85200880_D - Installation