Purge gas specification
Reference data
Maximum concentration of oil
0.1 parts per million
Permitted purge gas flow (when necessary) 20 to 50 sccm (0.33 to 0.84 mbar l s
-1
or
33 to 84 Pa l s
-1
)
Recommended purge gas flow
25 sccm (0.42 mbar l s
-1
, 42 Pa l s
-1
)
Maximum permitted supply pressure of
the purge gas
1 bar (gauge), 14.5 psig, 2 x 10
5
Pa
3.5 Cooling water
Refer to
Table: Cooling water specification on page 24
for the cooling water
specification. The cooling water supply must be a high
‑
quality drinking water
specification. Check with the water supply authority if there is doubt about the quality
of the supply.
Table 7
Cooling water specification
Cooling water specification
Reference data
Quality
Mechanically clean and optically clear
with no deposits or turbidity
pH value
6.0 to 8.0
Maximum calcium carbonate
concentration
75 parts per million
Maximum chloride concentration
100 parts per million
Minimum oxygen concentration
4 parts per million
Minimum cooling water flow rate (at
15 °C)
15 l hr
-1
Water temperature
10 to 20 °C
Maximum water pressure
5 bar (gauge), 73.5 psig, 6 x 10
5
Pa
Materials exposed to the cooling water
Nickel plated brass
3.6 Electrical data
The pumps can be operated by the customers system or by the manufacturer's TIC Turbo
Instrument Controller or TIC Turbo Controller.
If using the customer system, the size of the power supply depends on the application.
The rate of increase in the pump speed is dependent on the power limit setting. The
power limit setting overrides the necessary power supply setting. If the serial
communications or an access to the manufacturer's TIC is available, the power limit
setting of the pump can be selected. Refer to
Table: Logic interface technical data on
for the maximum power limit settings for the pumps. If the rapid cycling of the
pump is necessary for an application, you can get the faster rate of increase in the pump
speed if the power supply supplies higher current, up to a maximum in accordance with
Table: Logic interface technical data on page 27
.
Page 24
B85200880_D - Technical data