A541-11-880 Issue A
Page 50
© Edwards Limited 2010. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Appendix A2
Table A15 - Hexadecimal to digital conversion
BIT set combinations
Hexadecimal
Decimal
Combination
F
15
8+4+2+1
E
14
8+4+2
D
13
8+4+1
C
12
8+4
B
11
8+2+1
A
10
8+2
9
9
8+1
8
8
8
7
7
4+2+1
6
6
4+2
5
5
4+1
4
4
4
3
3
2 + 1
2
2
2
1
1
1
Table A16 - Inverter alarm codes
INVERTER ALARMS
BIT
set
Fault code
combination
PDT reported fault name
Description
1
8
ACCELERATION_TO
If motor has not accelerated to a minimum speed of 10Hz
within 60 secs then drive will try to restart 3 times before
giving alarm
4
OVERLOAD_TO
Alarm given when dry pump speed is below 7Hz for 3
minutes
2
SC_MODE_INTERLOCK
Internal fault contact Edwards
1
FLASH_DOWNLOAD_ FAULT
Internal fault contact Edwards
2
8
POST_FAULT
Internal fault contact Edwards
4
OSTEST _FAULT
Internal fault contact Edwards
2
EEPROM_FAULT
Internal fault contact Edwards
1
PWM_TRIP
Inverter output switched off, active when drive not
running. Restart pump or cycle power to reset.
3
8
MISSING_PHASE_ TIMEOUT
Indicates a missing input phase - check power connections
to pump and fuses in supply.
Warning should appear first and then dry pump trips after
10 minutes.
4
EMS
EMS has been activated.
Check EMS button and EMS plug in the back of pump.
2
DESAT_FAULT
Internal fault contact Edwards
1
UNDERT
Inverter is too cold. See warning register to determine
source of problem.