D395-90-880 Issue E
Page 24
© Edwards Limited 2010. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Appendix A
58
Incorrect pressure format
A command had a wrongly formatted pressure. Check format
specified in
.
81
No gauge connected
An attempt was made to read or control a gauge when it was
not connected. Check that gauge number is correct.
82
Unknown gauge type
A gauge is connected that cannot be recognised. Check that
the gauge is a type listed in
upgrade may be needed. See
83
Gauge not reading pressure
An attempt was made to read a gauge which was switched
off. Check that the gauge is switched on. Use the command
!GW if necessary.
84
AIM gauge striking
An attempt was made to read a gauge which was striking.
Wait until the gauge has struck and try again.
90
Incorrect gauge type, query/command
not appropriate
The command or query is not supported by the particular
gauge e.g. an attempt was made to switch an APG on or off.
91
Gauge turn-on is inhibited by link
A "GW1=1" command has been sent when 2 ADCs are linked,
and the pressure is too high to allow the controlled gauge to
turn on.
Table A1 - ADC error numbers (continued)
ERR
number
Meaning
Possible cause / remedy