![Edwards A37104902 Instruction Manual Download Page 17](http://html1.mh-extra.com/html/edwards/a37104902/a37104902_instruction-manual_2366076017.webp)
Motor output rating - continuous
0.16 kW
Motor electrical supply
Single phase
Table 5
Electrical data
Pump
Nominal supply (V)
Frequency (Hz)
Full load current (A)
E2M0.7
100 - 115
50
1.8
100 - 120
60
1.9
200 - 230
50
0.85
200 - 240
60
1.0
E2M1 and E2M1.5
100 - 115
50
3.1
100 - 120
60
3.0
200 - 230
50
1.3
200 - 230
60
1.4
Table 6
Recommended regional supply protection
Area
Voltage (V)
Rating (A)
UK
230
6
Europe
230
6
USA
110
10
Japan
100
10
3.5.1 Electrical cables
Recommended cord sets and fuses for regional requirements.
Table 7
Recommended cord sets
Description
Rating
Coupler type Item number
Cord set
assembly, UK
Cable style = H05VV-F, 3 x 1.0 mm
2
,
300 V,
70 °C, maximum length of 2.0 metres
Plug type = BS1363 UK plug
Appliance coupler = IEC60320 style C14
Fuse type = BS1363 10 Amp fuse, to an
IEC60320 style
Straight entry A50505000
Cord set
assembly,
Europe
Cable style = H05VV-F, 3 x 1.0 mm
2
,
300 V, 70 °C, maximum length of
2.0 metres
Plug type = European Schuko VDE
approved, 16 A 250 V rated with dual
earthing contact
Appliance coupler = IEC60320 style C14
Straight entry A50506000
Page 15
A37132880_F - Technical data
Summary of Contents for A37104902
Page 43: ...Page 41 A37132880_F Service...
Page 44: ...edwardsvacuum com...