Cause
The pump must be stopped before starting the solvent soak.
Remedy
Stop the pump and then repeat the command.
Fault
Event 331.42
Cannot start Command. The pump is not warmed up.
Cause
The pump must be warmed up and in the green mode/standby mode to initiate
the DP Clean sequence.
Remedy
- Either reduce the warm up temperature (refer to
on page 136), enable the pump on
‑
process when cold (refer to
on page 136) or wait for the pump to warm up.
- Make sure that the pump is in green mode/standby mode before attempting the
DP Clean.
Fault
Event 331.43
Cannot start Command. the pump is not in the green mode.
Cause
The pump must be in the green mode/standby mode to initiate the DP Clean
sequence.
Remedy
Set the pump into green mode/standby mode and then start the DP clean.
Fault
Event 331.46
Cannot start command. The pump is not running.
Cause
The pump must be running to initiate the DP Clean sequence.
Remedy
-
10.2 Warnings
The pump controller generates a warning when a problem is encountered.
By default, the pump is prevented from going on
‑
process when there are active
warnings so it is important to investigate the cause of the warning. It is possible to
configure the pump so that it will go on
‑
process with the active warnings. Refer to
a pump to go on-process with a warning
on page 137 for more information.
Once the problem that caused a warning has been resolved, the warning is cleared by
the pump controller.
10.2.1 LED warning indicators
If the pump encounters a problem, warnings are indicated on the LEDs on the front
control panel, the rear panel and on the PDT if fitted.
The warning LEDs on the front control panel and rear panel illuminate continuously
when a warning is generated.
Page 162
M58800880_H - Fault finding
Summary of Contents for 9773512584
Page 6: ...This page has been intentionally left blank ...
Page 87: ...Figure 19 GXS160 installation drawing Sheet 4 Page 83 M58800880_H Installation drawings ...
Page 178: ...This page has been intentionally left blank ...
Page 179: ...This page has been intentionally left blank ...
Page 180: ...edwardsvacuum com ...