> LOAD SAVED PROGRAM
MODIFY SAVED PROGRAM
CREATE NEW PROGRAM
DELETE SAVED PROGRAM
TEMP HOLD PARAMS
LOAD PROGRAM
>
330
336 341 669
331
337
701
000
332
952
004
000
333
961
371
111
334
962
764
000
PROGRAM 330
TEMP TIME
Init Denat
Denaturation
94 0040
Annealing
45 0045
Extension
72 0045
Final Ext
72 0300
Cycles
01 30
PROGRAM >000 TEMP TIME
Init Denat
00 0000
Denaturation
00 0000
Annealing
00 0000
Extension
00 0000
Final Ext
00 0000
Cycles
00 00
TEMPERATURE HOLD MENU
TEMP HOLD ON
> SET HOLD TEMP
010
RETURN
Quick Guide Screens
A
B
C
D
E
GETTING STARTED
Turn on the power switch (“|” mark is on, “O” mark is off).
Use the up or down button on the front panel to select the desired
function (the arrow to the left of the menu will allow you to choose
that selection).
The five options are as follows:
LOAD SAVED PROGRAM
MODIFY SAVED PROGRAM
CREATE NEW PROGRAM
DELETE SAVED PROGRAM
RUN PROGRAM
LOAD SAVED PROGRAM
Use this function if you are running
an EDVOTEK® kit. The programs for
most EDVOTEK® PCR kits are pre-
loaded on the EdvoCycler™ and
ready-to-run. Simply follow ED-
VOTEK instructions for preparing
DNA samples for PCR and place
tubes in the EdvoCycler™. The
block holds only 0.2 ml PCR tubes -
if PCR sample is prepared in a
different tube, you must transfer
sample to a 0.2 ml thin-wall PCR
tube (supplied in all EDVOTEK®
PCR kits).
1.
Scroll to
LOAD SAVED PROGRAM
and press enter.
2.
Scroll up or down to the desired program and press enter. Check
with the kit instructions to make sure parameters match with
screen display. If they do not, reference
MODIFY SAVED PRO-
GRAM
section and make the necessary changes to the existing
program.
continued
IMPORTANT
Always check the PCR cycling
parameters in the EDVOTEK®
manual that accompanies your
experiment. Compare the
program to ensure that it
matches the pre-programmed
cycles. If there has been a
change, it will be necessary to
modify the existing program to
match the new parameters.