AIR MOVEMENT WILL AFFECT THE TEMPERATURE READING.
WORK IN AN AREA WHERE THIS IS MINIMAL.
CALIBRATION
1. Using a clean Tip, turn on power
and set Temperature Control
Knob to approx. 500°F (260°C).
Allow Tip to warm up.
2. Tin the tip properly and place
center of the thermo-couple
wire on tip.
3. Apply a small amount of solder
on the center of the thermo-
couple wire, to form a good con-
tact between the tip and the
thermo-couple wire.
4. Set Temperature Control Knob
to 400°F (205°C).
5. Adjust
LO
-Temp. Calibration Pot
so the Meter will read 400°F
(205°C).
6. Set Temperature Control Knob
to 800°F (427°C).
7. Adjust
HI
-Temp. Calibration Pot
so the Meter will read 800°F
(427°C).
1. Set knob at 400°F.
2. Adjust LO to 400°F.
3. Set knob at 800°F.
4. Adjust HI to 800°F.
HI LO
Temperature
Control Knob
Calibration Pot