background image

VERIS

™ 

Service Manual

 

 

EMS & FMS Stimulators

 

VG-119-D / 21-Jan-2017 

 

Page 11 of 16 

10.

 

Stimulus Grid Calibration  (FMS Stimulators only

The retina does not reflect enough light to make it possible to see the stimulus in the IR fundus 
image

To make it possible for the operator to monitor where the subject is fixating on the multifocal 

stimulus array, Veris™ overlays an outline of the stimulus grid on the fundus image. This grid must 
be calibrated to match the actual stimulus. This is achieved using a model eye (Grid Calibrator) with 
a highly reflective fundus using the following procedure:

 

1.

 

With a test subject folder open, select the protocol containing the words ‘for grid calibration’. 
from the "FMS Microdisplay" Analysis Settings folder. 

2.

 

Open a recording window. 

3.

 

Place the Grid Calibrator on the FMS 
front lens. 

4.

 

Turn off IR illumination with the toggle 
switch at the back of the black 
microdisplay tube. 

5.

 

Select the fundus camera by toggling the up or down arrow keys of the computer keyboard. 

6.

 

Double-click on the fundus image in the window to open the camera controls. 

7.

 

Adjust the image brightness by means of and contrast controls and by rotating the Grid Calibrator 
as necessary. Fluctuations of image brightness are normal.  If you can still not see the stimulus 
well enough, you may open the lens aperture. To do so, move the camera back out of the FMS IV 
housing using the focusing knob until the black aperture lever is accessible. 

8.

 

Focus the video image 

9.

 

Open the "Stimulus" tab of the camera 
control window and adjust the size and 
position of the grid to match the stimulus as 
closely as possible.  (Holding the "

shift

key down on the keyboard while clicking 
on the size and position controls produces 
larger changes.) 

10.

 

Click OK to exit from the controls dialog, 
then exit from the recording window when 
the grid matches the stimulus as closely as 
possible.  Contact EDI if a satisfactory 
match cannot be achieved. 

11.

 

Reposition the CCD camera aperture to the 
position you normally use for fundus monitoring and switch the IR fundus illumination back on. 

This calibration will be correct for any multifocal stimulus array and for any refractive 
correction. 

EDI recommends repeating this procedure every 3-4 months or whenever a new version of the 
software has been installed. 

 

 

 

Summary of Contents for VERIS EMS IV

Page 1: ...ulators VG 119 D 21 Jan 2017 Page 1 of 16 VERIS FMS IV EMS IV Stimulators Installation Service Manual Release D Electro Diagnostic Imaging Inc http www veris edi com Tel 650 631 0120 Fax 650 631 0122...

Page 2: ...Information 3 1 4 Warnings and Precautions 3 2 Labels 5 2 1 Component Labels 5 2 2 Shipping Label 5 3 User Profile 6 4 Product Lifetime 6 5 Components of the FMS IV EMS IV Stimulators 6 6 Theory of O...

Page 3: ...Medical Electrical Equipment Part 2 General requirements for basic safety and essential performance collateral standard Electromagnetic VERIS has been tested for radiative emissions and shown to meet...

Page 4: ...he United States Canada and Europe If the VERIS System is used with customer provided components testing for safety is the responsibility of the customer Do not sit on the cart or table CAUTION Federa...

Page 5: ...as CA 95035 Made in USA Electro Diagnostic Imaging Inc 1551 McCarthy Blvd Milpitas CA 95035 Made in USA Electro Diagnostic Imaging Inc 1551 McCarthy Blvd Milpitas CA 95035 Made in USA Electro Diagnost...

Page 6: ...nob spherical refractive correction with a range of 20 to 8 diopters 7 Specifications Physical Stimulator unit Approx 18 21 x 7 x 4 25 46 54 cm x 19cm x 6 7 cm powder coated upon aluminum housing Note...

Page 7: ...tion monitoring during testing Software provides for capturing and saving single frames of the fundus image Recorded waveform traces later can be easily superimposed upon saved fundus image images Pat...

Page 8: ...ox against its packing slip when it is opened to account for all items Mount the arm on the right front edge of the desktop Mount the Control Box to the left of the Arm Mount Mount the tilting mechani...

Page 9: ...box labeled FMS Power Out The GearMo serial adapter connects directly to the control box serial port and to the serial port of the Grass 15LT amplifier If replacing a previously installed Keyspan Adap...

Page 10: ...m 5 Start VERIS open a protocol that uses the stimulator and select a plot 6 From Calibration on the menu bar select Calibrator Calibration and make sure Spot Calibrator is Checked 7 Cover the opening...

Page 11: ...dow to open the camera controls 7 Adjust the image brightness by means of and contrast controls and by rotating the Grid Calibrator as necessary Fluctuations of image brightness are normal If you can...

Page 12: ...e lens Apply lens cover when the instrument is not in use If necessary clean lens with eyeglass cleaning materials and soft cotton or micropore cloth 3 Calibration Perform Grid Calibration and Auto Ca...

Page 13: ...crodisplay control box Keyspan adapter USB port of computer Switch the microdisplay control box OFF wait then and ON again Then run VERIS and enter a recording window It may be necessary to try twice...

Page 14: ...le 76 6FT Cables to Go 1 Mini display DVI adapter Adapter 03 Apple Computers 1 Stimulator Arm Light IOP7500 500 124 EDI 1 Spot Calibrator CAL Spot 01 International Light EDI 1 Grid Calibrator CAL GRID...

Page 15: ...15 of 16 14 Cables Schematics Diagrams Calibrator DB 25 Connector Cable 76 02 and 76 6FT Standard Male Male VGA Legacy Cables Cable 76 01 Camera Interface Cable Serial Assembly Keyspan Adapter and Cab...

Page 16: ...VERIS Service Manual EMS FMS Stimulators VG 119 D 21 Jan 2017 Page 16 of 16 15 Software All stimulator functions are controlled by the VERIS application software...

Reviews: