9.

 

Stimulus Luminance Calibration  (FMS IV and EMS IV Stimulators)

 

Luminance calibration of the EMS IV/ FMS IV microdisplay requires a calibrated Spot Calibrator 
with Eye Camera Calibrator Adapter (ECCA). These components are provided with every Veris™ 
platform. 

 

 

 

 

 

 

 

Measuring the Calibrator Dark Current 

1.

 

Connect the FMS/EMS controller to the existing Veris system. 

2.

 

Power up the computer. 

3.

 

Turn on the power switch on the Controller. 

4.

 

Connect the calibrator’s 25-pin connector of the Spot 
Calibrator to the socket labeled “Auto-calibrator” on the 
VERIS interface box. A spot calibrator is provided with every 
Veris™ platform.

 

5.

 

Start 

VERIS

, open a protocol that uses the stimulator and 

select a plot.

  

6.

 

From Calibration on the menu bar select Calibrator Calibration 
and make sure Spot Calibrator is Checked 

7.

 

Cover the opening of the calibrator to keep the sensor in 
complete darkness. 

8.

 

Click on Read, wait until the the reading is reasonably stable, 
click stop and accept the reading of the dark current. 

9.

 

Slide the calibrator sensor with the Calibrator Adapter (ECCA) attached from the top over the 
front lens of the stimulator.

 

 

10.

 

From Calibration on the menu bar select Autocalibration, press start and wait until the process is 
finished. 

11.

 

The three curves shown must be monotonically increasing 
toward the right and look similar to those in the figure shown 
here. If this is verified, you may accept

 the calibration. 

 

 

Spot Calibrator with 

ECCA attached 

Spot Calibrator (right) with 

Eye Camera Calibrator Adapter (left)

 

Summary of Contents for VERIS EMS IV

Page 1: ...ulators VG 119 D 21 Jan 2017 Page 1 of 16 VERIS FMS IV EMS IV Stimulators Installation Service Manual Release D Electro Diagnostic Imaging Inc http www veris edi com Tel 650 631 0120 Fax 650 631 0122...

Page 2: ...Information 3 1 4 Warnings and Precautions 3 2 Labels 5 2 1 Component Labels 5 2 2 Shipping Label 5 3 User Profile 6 4 Product Lifetime 6 5 Components of the FMS IV EMS IV Stimulators 6 6 Theory of O...

Page 3: ...Medical Electrical Equipment Part 2 General requirements for basic safety and essential performance collateral standard Electromagnetic VERIS has been tested for radiative emissions and shown to meet...

Page 4: ...he United States Canada and Europe If the VERIS System is used with customer provided components testing for safety is the responsibility of the customer Do not sit on the cart or table CAUTION Federa...

Page 5: ...as CA 95035 Made in USA Electro Diagnostic Imaging Inc 1551 McCarthy Blvd Milpitas CA 95035 Made in USA Electro Diagnostic Imaging Inc 1551 McCarthy Blvd Milpitas CA 95035 Made in USA Electro Diagnost...

Page 6: ...nob spherical refractive correction with a range of 20 to 8 diopters 7 Specifications Physical Stimulator unit Approx 18 21 x 7 x 4 25 46 54 cm x 19cm x 6 7 cm powder coated upon aluminum housing Note...

Page 7: ...tion monitoring during testing Software provides for capturing and saving single frames of the fundus image Recorded waveform traces later can be easily superimposed upon saved fundus image images Pat...

Page 8: ...ox against its packing slip when it is opened to account for all items Mount the arm on the right front edge of the desktop Mount the Control Box to the left of the Arm Mount Mount the tilting mechani...

Page 9: ...box labeled FMS Power Out The GearMo serial adapter connects directly to the control box serial port and to the serial port of the Grass 15LT amplifier If replacing a previously installed Keyspan Adap...

Page 10: ...m 5 Start VERIS open a protocol that uses the stimulator and select a plot 6 From Calibration on the menu bar select Calibrator Calibration and make sure Spot Calibrator is Checked 7 Cover the opening...

Page 11: ...dow to open the camera controls 7 Adjust the image brightness by means of and contrast controls and by rotating the Grid Calibrator as necessary Fluctuations of image brightness are normal If you can...

Page 12: ...e lens Apply lens cover when the instrument is not in use If necessary clean lens with eyeglass cleaning materials and soft cotton or micropore cloth 3 Calibration Perform Grid Calibration and Auto Ca...

Page 13: ...crodisplay control box Keyspan adapter USB port of computer Switch the microdisplay control box OFF wait then and ON again Then run VERIS and enter a recording window It may be necessary to try twice...

Page 14: ...le 76 6FT Cables to Go 1 Mini display DVI adapter Adapter 03 Apple Computers 1 Stimulator Arm Light IOP7500 500 124 EDI 1 Spot Calibrator CAL Spot 01 International Light EDI 1 Grid Calibrator CAL GRID...

Page 15: ...15 of 16 14 Cables Schematics Diagrams Calibrator DB 25 Connector Cable 76 02 and 76 6FT Standard Male Male VGA Legacy Cables Cable 76 01 Camera Interface Cable Serial Assembly Keyspan Adapter and Cab...

Page 16: ...VERIS Service Manual EMS FMS Stimulators VG 119 D 21 Jan 2017 Page 16 of 16 15 Software All stimulator functions are controlled by the VERIS application software...

Reviews: