background image

VERIS

™ 

Service Manual

 

 

EMS & FMS Stimulators

 

VG-119-D / 21-Jan-2017 

 

Page 1 of 16 

 
 
 
 

VERIS™ 

FMS IV / EMS IV Stimulators 

 
 
 

Installation & 

Service Manual 

Release D 

 
 
 
 

 
 

 

 
 

 
 

 
 

 
 

 
 

 

 

Electro-Diagnostic Imaging, Inc. 
http://www.veris-edi.com 

Tel  650-631-0120 
Fax  650-631-0122 
Email: [email protected]

 

0413

Summary of Contents for VERIS EMS IV

Page 1: ...ulators VG 119 D 21 Jan 2017 Page 1 of 16 VERIS FMS IV EMS IV Stimulators Installation Service Manual Release D Electro Diagnostic Imaging Inc http www veris edi com Tel 650 631 0120 Fax 650 631 0122...

Page 2: ...Information 3 1 4 Warnings and Precautions 3 2 Labels 5 2 1 Component Labels 5 2 2 Shipping Label 5 3 User Profile 6 4 Product Lifetime 6 5 Components of the FMS IV EMS IV Stimulators 6 6 Theory of O...

Page 3: ...Medical Electrical Equipment Part 2 General requirements for basic safety and essential performance collateral standard Electromagnetic VERIS has been tested for radiative emissions and shown to meet...

Page 4: ...he United States Canada and Europe If the VERIS System is used with customer provided components testing for safety is the responsibility of the customer Do not sit on the cart or table CAUTION Federa...

Page 5: ...as CA 95035 Made in USA Electro Diagnostic Imaging Inc 1551 McCarthy Blvd Milpitas CA 95035 Made in USA Electro Diagnostic Imaging Inc 1551 McCarthy Blvd Milpitas CA 95035 Made in USA Electro Diagnost...

Page 6: ...nob spherical refractive correction with a range of 20 to 8 diopters 7 Specifications Physical Stimulator unit Approx 18 21 x 7 x 4 25 46 54 cm x 19cm x 6 7 cm powder coated upon aluminum housing Note...

Page 7: ...tion monitoring during testing Software provides for capturing and saving single frames of the fundus image Recorded waveform traces later can be easily superimposed upon saved fundus image images Pat...

Page 8: ...ox against its packing slip when it is opened to account for all items Mount the arm on the right front edge of the desktop Mount the Control Box to the left of the Arm Mount Mount the tilting mechani...

Page 9: ...box labeled FMS Power Out The GearMo serial adapter connects directly to the control box serial port and to the serial port of the Grass 15LT amplifier If replacing a previously installed Keyspan Adap...

Page 10: ...m 5 Start VERIS open a protocol that uses the stimulator and select a plot 6 From Calibration on the menu bar select Calibrator Calibration and make sure Spot Calibrator is Checked 7 Cover the opening...

Page 11: ...dow to open the camera controls 7 Adjust the image brightness by means of and contrast controls and by rotating the Grid Calibrator as necessary Fluctuations of image brightness are normal If you can...

Page 12: ...e lens Apply lens cover when the instrument is not in use If necessary clean lens with eyeglass cleaning materials and soft cotton or micropore cloth 3 Calibration Perform Grid Calibration and Auto Ca...

Page 13: ...crodisplay control box Keyspan adapter USB port of computer Switch the microdisplay control box OFF wait then and ON again Then run VERIS and enter a recording window It may be necessary to try twice...

Page 14: ...le 76 6FT Cables to Go 1 Mini display DVI adapter Adapter 03 Apple Computers 1 Stimulator Arm Light IOP7500 500 124 EDI 1 Spot Calibrator CAL Spot 01 International Light EDI 1 Grid Calibrator CAL GRID...

Page 15: ...15 of 16 14 Cables Schematics Diagrams Calibrator DB 25 Connector Cable 76 02 and 76 6FT Standard Male Male VGA Legacy Cables Cable 76 01 Camera Interface Cable Serial Assembly Keyspan Adapter and Cab...

Page 16: ...VERIS Service Manual EMS FMS Stimulators VG 119 D 21 Jan 2017 Page 16 of 16 15 Software All stimulator functions are controlled by the VERIS application software...

Reviews: