Calibration consists in a standard battery charge followed by a deep discharge, and a
complete charge. This procedure usually takes between 10 and 13 hours, whereas a standard
charge only takes about 3.5 hours.
Calibrate batteries by placing them in the optional, external charger and then pressing the
recalibrate button. (See Charging the Batteries on page 35)
Eddyfi recommends calibrating your batteries at least once every six months.
During calibration, the charger may become warm.
3.
Troubleshooting
To troubleshoot the Ectane 3, you must first connect it to a workstation running Magnifi®.
Troubleshooting information is found in the Magnifi documentation.
4.
Firmware Corruption
If the Ectane 3 firmware becomes corrupted, you must start the instrument from a backup
firmware. This procedure is explained in the Magnifi documentation.
To locate the backup firmware button, see Battery Compartment on page 23.
Summary of Contents for Ectane 3
Page 1: ...Eddyfi Ectane 3 User s Manual Version 1 0 Date December 2022...
Page 6: ...6 Chapter 1 Important Information Safety Precautions and Conventions...
Page 14: ...14 Chapter 2 Introducing Ectane...
Page 18: ...18 Figure 4 2 Ectane 3 E64 and E128 Figure 4 3 Ectane 3 E256...
Page 19: ...19 Figure 4 4 Ectane 3 I Figure 4 5 Ectane 3 ERNM...
Page 20: ...20 Figure 4 6 Ectane 3 ERNMI Figure 4 7 Ectane 3 E64RNM and E128RNM...
Page 21: ...21 Figure 4 8 Ectane 3 E64RNMI and E128RNMI...
Page 25: ...25 Chapter 3 Preparing the Instrument...
Page 34: ...34 Chapter 4 Managing Batteries...
Page 39: ...39 Chapter 5 Connector References...
Page 51: ...51 Chapter 6 Specifications...
Page 59: ...59 Chapter 7 Maintenance Troubleshooting...
Page 62: ...62 Appendix A Accessories...
Page 64: ...64 Appendix B Legal Notice...