47
Pin
I/0
Signal
Description
H
Output
Alarm
Used to indicate that the probe is in the
air (for tube inspections) (open drain with
10 kΩ pull
-up).
J
Relay
contact
Relay 2
Relay used for automatic acquisition
sequence control
K
Relay
contact
Relay 3
Relay used for automatic acquisition
sequence control
L
Relay
contact
Relay 1
Relay used for automatic acquisition
sequence control
M
TTL Input
CLKACQ
Receives signal to trigger acquisition with
probe position along the scanning axis
N
NC
No connection - Reserved
P
TTL Input
Φ
B2
Phase B axis 2
R
TTL Input
Φ
A3
Input 5
Phase A axis 3
User-defined input for automatic
acquisition sequence
S
Supply
Output
+5 V
T
TTL Input
Φ
B3
Input 4
Phase B axis 3
User-defined input for automatic
acquisition sequence
U
TTL Input
Φ
A3
Phase A axis 3
Summary of Contents for Ectane 3
Page 1: ...Eddyfi Ectane 3 User s Manual Version 1 0 Date December 2022...
Page 6: ...6 Chapter 1 Important Information Safety Precautions and Conventions...
Page 14: ...14 Chapter 2 Introducing Ectane...
Page 18: ...18 Figure 4 2 Ectane 3 E64 and E128 Figure 4 3 Ectane 3 E256...
Page 19: ...19 Figure 4 4 Ectane 3 I Figure 4 5 Ectane 3 ERNM...
Page 20: ...20 Figure 4 6 Ectane 3 ERNMI Figure 4 7 Ectane 3 E64RNM and E128RNM...
Page 21: ...21 Figure 4 8 Ectane 3 E64RNMI and E128RNMI...
Page 25: ...25 Chapter 3 Preparing the Instrument...
Page 34: ...34 Chapter 4 Managing Batteries...
Page 39: ...39 Chapter 5 Connector References...
Page 51: ...51 Chapter 6 Specifications...
Page 59: ...59 Chapter 7 Maintenance Troubleshooting...
Page 62: ...62 Appendix A Accessories...
Page 64: ...64 Appendix B Legal Notice...