default value after the oximeter is turned on again.
Press the
Up Arrow
or
Down Arrow
button to
increase or decrease values.
Choose
SpO
2
Lo Alarm
in the menu, press the
Function
button to set up it. The
SpO
2
Lo Alarm
box
will change from real line box to broken line box. The
adjustable range for the lower limit of SpO
2
Alarm is
from 0 to “the upper limit of SpO
2
Alarm - 1”. If the
value of SpO
2
Lo Alarm is set to less than 85, it will
restore to 85 after the oximeter is turned on again.
Press the
Up Arrow
or
Down Arrow
button to
increase or decrease values.
SpO
2
Hi Alarm
is always higher than
SpO
2
Lo Alarm
by at least 1%.
Press the
Function
button, confirm the alarm range
setup.
Press the
On/Off
button to exit the menu, and return to
measurement state.
3.3.11 PR Alarm Setup
The user can use
PR Hi Alarm
and
PR Lo Alarm
in menu to
adjust pulse rate alarm limits.
By default,
PR Hi Alarm
and
PR Lo Alarm
are 170 and 40
respectively.