VE-H100B Veterinary Pulse Oximeter User Manual
- 87 -
RF emissions
CISPR11
Class B
VE-H100B and charger
stand is suitable for use in
all establishments other
than domestic and those
directly connected to the
public low-voltage power
supply net work that
supplies buildings used for
domestic purpose.
Harmonic
emissions
IEC/EN61000-3-2
Pulse
oximeter: N/A
Charger
stand :Class A
Voltage
fluctuations
/flicker emissions
IEC/EN61000-3-3
Pulse
oximeter: N/A
Charger stand:
Complies
A2.2 Electromagnetic Immunity
For pulse oximeter:
Guidance and manufacturer’s declaration – electromagnetic
immunity
VE-H100B
is intended for use in the electromagnetic environment
specified below. The customer or the user of VE-H100B should
assure that it is used in such an environment.
Emissions test
Compliance Compliance
level
Electromagneti
c environment
- guidance
±6kV
contact
±6kV contact
±8kV air
Floors should
be wood,
理邦保密文件
EDAN CONFIDENTIAL