SE-1200 Series Electrocardiograph Service Manual Electrocardiograph Disassembly
- 52 -
4)
SE-1200 Express Display Screen Assembly
The following figure shows SE-1200 Express display screen assembly.
⑩
and
are sponge strips used to avoid direct touch among the display screen frame
,
upper casing
⑤
and LCD screen
.
is a plastic gasket.
is a hexangular nut.
5)
Disassembly Procedures of the Lower Unit (SE-1200 Express/SE-1200)
In the following figure,
①
is the lower unit, and
②
is a pan head screw securing the upper and
the lower unit. Remove all the related cables that can be moved out directly, and dismantle the
unit in the following sequence: ECG board
⑤
→ power supply socket components
⑥
(including
gasket
⑦
and grounded bar
⑧
) → the remainder cables → main board
④
→ paper tray
component of the recorder
③
→ PS900D power supply board
⑨
→ recorder frame plate
.