SE-1200 Series Electrocardiograph Service Manual Safety Guidance
- 6 -
WARNING
12. Electrodes of dissimilar metals should not be used; otherwise it may cause a high
polarization voltage.
13. The disposable electrodes can only be used for one time.
14. Do not touch the patient, bed, table or the equipment while using the ECG together
with a defibrillator.
15. Do not touch accessible parts of non-medical electrical equipment and the patient
simultaneously.
16. The use of equipment that applies high frequency voltages to the patient (including
electrosurgical equipment and some respiration transducers) is not supported and
may produce undesired results. Disconnect the patient data cable from the
electrocardiograph, or detach the leads from the patient prior to performing any
procedure that uses high frequency surgical equipment.
17. If the wireless AP technology is used, in order to maintain compliance with the FCC
RF exposure guidelines, the wireless AP should be installed and operated with a
minimum distance of 20cm between the radiator and the human body. There should
be no shield in or around the room where the wireless AP is used.
18. Fix attention on the examination to avoid missing important ECG waves.
19.
SHOCK HAZARD
-
Don’t connect non-medical electrical equipment, which has been
supplied as a part of the system, directly to the wall outlet when the non-medical
equipment is intended to be supplied by a multiple portable socket-outlet with an
isolation transformer.
20.
SHOCK HAZARD
-
Don’t connect electrical equipment, which has not been supplied
as a part of the system, to the multiple portable socket-outlet supplying the system.
21. Do not connect any equipment or accessories that are not approved by the
manufacturer or that are not IEC/EN 60601-1 approved to the electrocardiograph. The
operation or use of non-approved equipment or accessories with the
electrocardiograph is not tested or supported, and electrocardiograph operation and
safety are not guaranteed.
22. Any non-medical equipment (such as the external printer) is not allowed to be used
within the patient vicinity (1.5m/6ft.).