SE-1200 Series Electrocardiograph Service Manual Inspection and Troubleshooting
- 37 -
6.
Battery Test
On the
System Test
screen, press
Tab
or
Shift + Tab
, or the Up, Down, Left or Right arrow to
move the cursor to the
Battery
item, and then press
Enter
to display the
Battery
screen.
This screen displays the battery capacity, whether AC power is used and whether the battery is
being charged.
Press
Esc
to exit the screen.
Figure 5-10 Battery Screen
7.
Recorder Test
On the
System Test
screen, press
Tab
or
Shift + Tab
, or the Up, Down, Left or Right arrow to
move the cursor to the
Recorder
item, and then press
Enter
to display the
Recorder
screen.
On this screen, press the
PRINT/STOP
key to begin to print the triangle waves in effective paper
width. The status of the print head can be estimated from the triangle waves. Press
PRINT/STOP
again to stop printing.
Press
Esc
to exit the screen.
NOTE
: During the printing course, you should not press the
Esc
key to exit.