FTS-3 Fetal Telemetry System Service Manual
Periodic Tests
- 41 -
3)
Set the MHR as 60 bpm, 80 bpm, 100 bpm, 120 bpm and 140 bpm in turn.
4)
Check if the value displayed on the monitor screen is accurate. Error within ± 2 bpm is
acceptable.
11.2.4 DECG Performance Test
This test checks the DECG performance of the TOCO-E transducer.
Required equipment: FLUKE PS320 simulator.
1)
Connect the FTS-3 base station to the monitor and switch on them.
2)
Take the TOCO-E transducer from the base station and the transducer will switvh on.
Use the DECG cable to connect the transducer and the simulator as shown below.
Figure 11-5
3)
Set the FHR as 60 bpm, 90 bpm, 120 bpm, 150 bpm, 180 bpm and 210 bpm in turn.
4)
Check if the value displayed on the monitor screen is accurate. Error within ± 2 bpm is
acceptable.