FTS-3 Fetal Telemetry System Service Manual
Disassembling FTS-3
- 28 -
Chapter 10 Disassembling FTS-3
WARNING
1 Only qualified service personnel should open the base station.
2 Switch off the base station and remove the AC power cable before disassembling the
base station.
3 After any repair of the base station, perform safety tests before using the base staion
again.
10.1 Tools Required
Figure 10- 1
10.2 Blown Fuses
WARNING
Switch off the base station and remove the power cord before changing the fuse.
Replace the fuse when it is blown.
The two fuses of the base station are located on the rear panel, their specifications are:
Size: Ф5mm*20mm; Model: T2AH250V.
To replace a fuse:
1)
Place the base station on a flat surface and remove the power cord.
2)
Reverse the base station and pull the fuse container out as far as it can go.
1 - A cross-head screwdriver
2 - A pair of pliers
3 –A pair of
tweezers