Disassembling FTS-3
- 32 -
Figure 10- 8
No. Part
1
Upper Assembly
2
Power connecting line to indicative light board
3
Lower Assembly
4
Main board connecting line to indicative light board
(7)
Disconnect the cables/wires to separate the two parts.
NOTE:
The interfaces of the connecting line are fixed with the melt adhesive. Please drop alchol
and remove the melt adhesive with a pair of pliers. And the remove the pull out the
connecting line.