SE-3 3-Channel Electrocardiograph User Manual
-3-
conducting objects.
13. Electrodes with defibrillator protection should be used while defibrillating.
14. There is no danger for patients with pacemaker. However, if a pacemaker is
used, the results given by the equipment may be invalid, or lose the clinical
significance.
15. Do not touch the patient, bed, table and the equipment while using defibrillator
or pacemaker simultaneously.
16. In order to avoid burning, please keep the electrode far away from the radio
knife while using electrosurgical equipment simultaneously.
17. Always use electrode gel with reusable electrodes during defibrillation as ECG
recovery will be greater than 10 seconds. EDAN recommends the use of
disposable electrodes at all times.
WARNING
:
18. Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC/EN standards (e.g. IEC/EN 60950 for
data processing equipment and IEC/EN 60601-1 for medical equipment).
Furthermore all configurations shall comply with the valid version of the
standard IEC/EN 60601-1-1. Therefore anybody, who connects additional
equipment to the signal input connector or output connector to configure a
medical system, must make sure that it complies with the requirements of the
valid version of the system standard IEC/EN 60601-1-1. If in doubt, consult our
technical service department or your local distributor.
19. The summation of leakage current should never exceed leakage current limits
while several other units are used at the same time.
20. The potential equalization conductor can be connected to that of other
equipment when necessary, to make sure that all these equipment are
connected with the potential equalization bus bar of the electrical installation.
1.2.2 Battery Care Warnings
WARNING
:
21. Improper operation may cause the battery to be hot, ignited or exploded, and it
may lead to the declination of battery
’
s capacity. It is necessary to read the user