8.2 Disinfection
To avoid permanent damage to the equipment, disinfection can be performed only when it has
been considered as necessary according to your hospital
’
s regulations.
Before disinfection clean the equipment first. Then wipe the surface of the unit and patient
cable with hospital standard disinfectant.
CAUTION
:
Do not use chloric disinfectant such as chloride and sodium hypochlorite etc.
8.3 Care and Maintenance
8.3.1 Recharge and Replacement of Battery
1) Capacity Identification
Current capacity of the rechargeable battery can be identified according to the battery
symbol in the top right corner on LCD screen.
: Full capacity
: Capacity is limited, and recharge should be taken into account
: Battery is weak; and warning message
“
BAT WEAK
”
will be displayed on LCD screen.
The battery should be recharged immediately
2) Recharge
SE-3 is equipped with recharge control circuit together with built-in rechargeable lithium
battery. When connect with the mains supply, the battery will be recharged automatically.
And then the battery recharge indicator lamp (
) and the mains supply indicator lamp (
)
will be lit at the same time. During the course of recharging, the symbol
“
”
will flash in
the top right corner of LCD screen. When the capacity of battery is full, the symbol
“
”
will stop flashing, and the battery recharge indicator lamp (
) will usually be black. But
if SE-3 is power off, the lamp will still lit just because the equipment will not monitor the
recharge status; so you need to power on the device to verify the status.
Because of the capacity consumption during storage and transport, the capacity of battery is
not full while using at the first time. Battery recharge should be considered before first