VE-300 Veterinary Electrocardiograph User Manual
-51-
DFT Filter: 0.05/0.15/0.25/0.5
EMG Filter: 25Hz/35Hz/45Hz/OFF
LOWPASS Filter:150Hz/100Hz/75Hz
CMRR:
≥
110dB
Sampling Frequency:
1000Hz
Patient Leakage Current:
<10
µ
A (220V~240V)
Patient Auxiliary Current:
<0.1
µ
A (DC)
Dielectric Strength:
4000V rms
Input:
≥
100k
Ω
; Sensitivity 10mm/V
±
5%;
Single ended
External
Input/Output
(Optional)
Output:
≤
100
Ω
; Sensitivity 1V/mV
±
5%;
Single ended