VE-300 Veterinary Electrocardiograph User Manual
-50-
Charge voltage (standard) = (17
±
0.1V)
Cycle life
≥
300 times
Power Consumption:
35VA (max)
Fuse:
T400mA250V
Ø
5
×
20/T200mA 250V
Ø
5
×
20
Recorder:
Thermal dot-matrix printer
Record Paper:
Folded thermal paper, 80mm width
Rolled thermal paper, 80mm width
Effective Width:
72mm
Paper Speed:
5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s,
50mm/s
(
±
3%
)
Recording
Accuracy of data:
±
5% (x-axis),
±
5%(y-axis)
Technique:
Peak-peak detection
HR Range:
30 BPM ~300 BPM
HR
Recognition
Accuracy:
±
1BPM
Leads:
7 standard leads
Acquisition Mode:
simultaneously 7 leads
A/D Resolution:
12 bits
Time Constant:
≥
3.2s
Frequency Response:
0.05Hz ~ 150Hz(-3dB)
Sensitivity:
2.5, 5, 10, 20 (mm/mV)
Input Impedance:
50M
Ω
(10Hz)
Input Circuit Current:
≤
50nA
Input Voltage Range:
<
±
5 mVpp
Calibration Voltage:
1mV
±
3%
DC Offset Voltage:
±
500mV
Noise:
<12.5
µ
Vp-p
Multichannel crosstalk:
≤
0.5mm
ECG Unit
Filter:
AC Filter: On/Off