ALIGNMENT
Setup
Connect the Model 441A as shown in Figure 6-3 and
set S1 to On.
Screen Adjustment
From the
OPERATE
screen, press and hold the dis-
play/push button for one second to activate the
VIEW ANGLE
screen. Adjust the encoder counter-
clockwise and clockwise to verify the view angle is
adjustable in both directions. Turning it clockwise dark-
ens the screen and counterclockwise lightens it. When
finished, set the knob for the best viewing contrast.
Voltage Alignment
While still at
VIEW ANGLE
, press the display once
to go to
ADJUST
−
10 V
. Adjust the encoder until
−
10.000 V
±
0.005 V is obtained on the DMM. The
three digits on this display (maximum range of
±
750
counts) are provided as a reference if needed. The least
count is
≈
1.3 mV. Record the DMM reading.
Press the display again to go to
ADJUST
+
10 V
. Rotate
the encoder until the DMM reads
+
10.000 V
±
0.005 V.
The three digits on this display (maximum range of
±
750 counts) are provided as a reference if needed. The
least count is
≈
1.3 mV. Record the DMM reading.
Press the display again to the return to the
OPERATE
screen.
0RGHO
$
2Q
Figure 6-3
Alignment
Alignment and Calibration
Model 441A
6-4
Summary of Contents for 441A
Page 18: ...Figure 3 1 Operational state Diagram Operation Model 441A 3 2...
Page 46: ...Applications Model 441A 4 16...
Page 50: ...Theory of Operation Model 441A 5 4...
Page 58: ...Alignment and Calibration Model 441A 6 8...
Page 62: ...Parts Lists Model 441A 7 4...
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 68: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 74: ......