90
5 TROUBLESHOOTING AND MESSAGES
The “
Detector damaged
” message appears.
There is a problem with the “VIRTUOSO” detector. Switch the detector off,
and then turn it on. If the message appears again, contact the service center or the
manufacturer.
The “
Could not calibrate detector. Potassium was not found
” message
appears during the energy calibration.
Try to change the detector position (for example, place the detector as close
to a wall of the room as possible or use the potassium chloride fertilizer of 1 kg
weight) and repeat a calibration.
The “
Parameter out of range
” message appears during the calibration
relative to energy.
Tap the “Restore coefficients” button in the energy calibration window to
restore coefficients. Repeat calibration following the given recommendations on
preparing for a calibration (see Section 4 “Energy calibration”). If the message
appears again, contact a manufacturer.