15
1.6 Charging the storage battery of the detecting device
The external battery charger, which is included in the radiometer’s
kit, is
connected to the connector (9). When you connect the charger to the detecting
device,
the “BC” LED
(meaning that external power is supplied) and the
"BATTERY" LED should light up. Upon completion of the charging process, the
"BATTERY" LED should go out and the "BC" LED should stay on. The "BC"
LED goes out only after removing the charger from the detecting device.
The process of battery charging does not affect the performance of the
device. If necessary, the detecting device can continuously operate from an external
power supply.
Important!
Charge the battery of the detecting device within the ambient
temperature range from 0 to 45
оС. Noncompliance with this temperature range can
cause failure or significant reduction in capacity of Li-Po battery.
Important!
In case of the storage battery (11) failure it should be replaced
with a new one of the same type (14500 (AA) Li-Po battery). Unscrew the screws
(12), remove the lid (4) and insert a new battery in the battery compartment (10)
observing the polarity. Replacement of a faulty battery can be performed by the
manufacturing company of the detecting device or its authorized service center.
Important!
If Li-Po battery is improperly installed, the radiometer goes out
of service. Warranty does not apply to this damage and the client is solely
responsible for the failure.
Summary of Contents for VIRTUOSO RKG-14
Page 1: ...RKG 14 VIRTUOSO multipurpose environment activity radiometer Operating manual I 412139 003 02...
Page 2: ......
Page 4: ...2 Version 8 Last updated June 2021...
Page 76: ...74 a b Figure 2 59 Track recording Figure 2 60 Track pausing...
Page 98: ...96 b VIRTUOSO measurement report...
Page 99: ...97 VIRTUOSO measurement report...