ecom-EN2
After sufficient purging time (X behinds CO disappears) the sensor
can be reintegrated into the measurement system with <F3> (if you
choose „Yes“ at „Adjustments“ / „Internal“ / „CO-Automatic“ the
CO sensor switches to measurement automatically). The sensor can
also be manually be excluded from the measurement system with
<F3>.
Probe and tubing
Depending on the frequency of use, probe and tubing should be
regularly cleaned in order to release particle deposits and to prevent
early wearing due to corrosion.
-Release the connections at the instrument and at the probe grip
to free the tubing.
-Clean it (flow warm water in then dry respectively blow water drops
out.
-Slightly grease the O-rings of the tubing connections from time to
time with acid-free grease.
Change printer paper roll
-Release the printer cover (press lock downwards).
-If necessary, extract the paper rest out of the printer. Hereto select
"Adjustments"/"Paper feed"/<Enter>).
-Remove the plastics tube of the previous roll.
-Insert the paper end in the slot under the transport roll (paper
roll inner side facing you while inserting the paper).
-Convey approx. 3 cm paper thru the printer ("Adjustments"/"Paper
feed"/<Enter>).
-Lay the paper roll in the corresponding hollow.
-Insert the paper thru the slot of the printer compartment cover
and close the latest.
Locking
printer cover