ecom-EN2
Page 39
The gas losses measurement is to be
performed with the instrument´s sam-
pling probe in the gas core stream after
menu call up (see chapter 8.1.). Once
the measurement is recorded with
<Memory> (disk symbol) the conver-
sion of the measurement results in neg-
ative points is available under the menu
point „Results“.
The surface losses measurement is
performed by a temperature sensor
specific for surfaces. The temperature
difference between boiler surface and
room temperature (air temperature sen-
sor) is determined and the percentage
loss is calculated. Once the menu point
is called up, the boiler performance
must be inputted. To easy the meas-
urement width, depth and height of the
boiler can also be entered (dimensions
will be memorized for surface calcula-
tion). Please proceed as follows:
- activate respective input window
with <Enter>
- inputs values using the keys numerical
function
- confirm input with <Enter>
or:
- adjust values using the cursor
keys <Right/Left>
If no boiler dimensions are inputted,
press <F1> (Start) to activate the
measurement recording. Here the di-
mensions for all surfaces must be in-
putted.
Measurement stored in
intermediate memory
O2
3.2 %
CO2
13.1 %
CO
0 ppm
Eff.
92.5 %
Losses
7.5 %
Exc. air
1.18
T.Gas
184 °C
T.Air
20 °C
BImSchV
Temperature sensor
for surfaces
P.Boiler 24.5 kW
Width 1.20 m
Depth 1.20 m
Hight 1.20 m
Modify:
↑↓< >
↑↓< >
↑↓< >
↑↓< >
↵
Surface loss
Start
+
Start