ecom-EN2
Page 33
DRT <-> PC!:
Load data:
Enables the data import from e.g. ecom software (available on our
website „www.ecom.de“). See chapter „Technical Data“ for data for-
mat information (please observe the transfer options of your soft-
ware!).
Proceed as follows:
-Connect ecom-EN2 and PC via USB cable.
-Select “Load data“ and confirm with <Enter>.
-Answer the displayed question with YES (<F1>).
-Decide if the data recorded can be cancelled
(<F1> for YES / <F4> for NO).
-Start the data transfer on your PC.
Send data:
With this function the data records completed with measurement
values can be transferred to the PC program (procedure similar to
chapter „Load data“).
12.2. Data logger
Here a Data logger record (“Data logger” and key <OK> / on top
right on the display appears the disk symbol) can be started or fin-
ished (just available when using the multi-media card). With <Rec-
ord> you can interrupt and continue the record (only in gas analysis).
For each recording one file will be written on the card. The files will
be numbered consecutively (J2KDL-00.csv, J2KDL-01.csv and so
on) and can be transferred to PC with a card reader. The length of a
dataset is 500 byte which means that on a 32 MB card 64000 meas-
urements could be recorded.
Save to MMC (min. 1 second / max. 999 seconds):
Adjustment of the interval time for data logger recordings.
CSV+Header (<F1> for YES / <F4> for NO):
Adjustment of data logging with or without column headings.