Page 28
ECOM-D
DRT <-> PC!:
Load data:
Enables data importation from the rbr software (available on our web-
site <www.rbr.de>). See the chapter on <Technical Data> for data
format information (please observe the transfer options of your soft-
ware!).
Proceed as follows:
-Connect ECOM-D to your PC via USB cable.
-Select
<Load data>
and confirm with
<Enter>
.
-Answer the displayed question with
YES
(
<F1>
).
-Decide if the data recorded can be cancelled
(
<F1>
for
YES
/
<F4>
for
NO
).
-Start the data transfer to your PC.
Send data:
With this function, the data records completed with measurement val-
ues can be transferred to the PC.
(the procedure is similar to
<Load data>
).
Data logger (Option):
Here a Data logger record (Select
<Data logger>
and press
<OK>
/
On the top right of the display, the disk symbol appears.) can be start-
ed or finished (only available when using the multi-media card). With
<Record>,
you can interrupt or continue the record (only in gas anal-
ysis). For each recording, one file will be written on the card. The files
will be numbered consecutively (J2KDL-00.csv, J2KDL-01.csv and so
on) and can be transferred to your PC with a card reader. The maxi-
mum size of a dataset is 500 bytes which means that on a 32 MB card
64000 measurements can be recorded.
In addition to data logger recordings, the data can also be transferred
with a USB cable (
USB Driver
/
1200 Baud
/ Protocol
DAS
) to the
software
“DASNT”
. The software
“DASNT”
and the
USB Driver
are
available free of charge from the rbr website.
Save to MMC (min. 1 second / max. 255 seconds):
Adjustment of the interval time for data logger recordings.