4.3
Site Services
The
STORM™
unit requires the following site services;
Service
Specification
Details
Power
90-264VAC, 50/60Hz, 6 A (max.)
(or) 180-550VAC, 50/60Hz
(or) 24VDC
System power and control
Water
20 l/min @ 3 bar (peak),
0.5l/min (consumed)
Tube Cleaning
–
potable supply
recommended.
Air
10 l/min @ 6 bar
Vacuum generation, Cylinder purging.
4.4
Electrical Terminations
Power
The STORM™ can accept 90
-264VAC single phase supply as standard.
Alternatively, 180-550VAC two phase, or 24VDC options can be supplied
–
to be specified at
order.
Input voltage range
Power Terminations
Standard
90-264VAC
L1 (Active)
N (Neutral)
Earth
Optional
180-550VAC
L1 (Phase A)
L2 (Phase B) Earth
Optional
24VDC
24 (24VDC+)
0 (0VDC)-
Earth
Figure 4-2 : Power Terminations