![Ecolab FAS1543 User Manual Download Page 17](http://html1.mh-extra.com/html/ecolab/fas1543/fas1543_user-manual_2354496017.webp)
Patient Safeguard System Installation Tool Directions for Use Ecolab, Inc.
Page 15
Figure 15
–
Display in Tool Info Mode (First Screen), System B
Figure 16
–
Display in Tool Info Mode (Second Screen), System B
Selecting Tool Info mode will allow the viewing of information about the Tool. The f
irst screen displays the Tool’s
firmware revision and battery level as a percentage. Screens two through four give information about the tuning
of the antennas for the Tool’s 125 KHz receiver. Pressing the Menu
key will return the Tool to the main menu.
5.12 Replacing the Batteries
Open the battery compartment on the back of the Tool by sliding the battery door towards the bottom end of the
Tool. Remove both old AA alkaline batteries from the battery compartment and discard. Place two new AA
alkaline batteries in the battery compartment following the diagram inside the battery compartment. Replace the
battery door back on the Tool by sliding the door towards the middle of the Tool.
6.0 Troubleshooting
6.1 The T
ool’s d
isplay will not turn on when the Select key is pressed
A. Ensure that the Power Switch is pressed down on the right side.
B. Replace the two AA alkaline batteries.
6.2 Beacon events do not appear on the dashboard
A. Ensure that the Tool is in the Bentley mode. The Beacon will not send an event over the network if the
Tool is in any other mode.
B. Ensure that the Beacon is working. This can be done by putting the Tool in Beacon mode and testing the
Beacon.
C. Ensure that the network is working. This can be done by putting the Tool in Network mode and testing
the n
etwork connection at the Beacon’s location.
7.0 Cleaning the Tool
The Tool should be cleaned by wiping with a soft cloth. A mixture of 90% water and 10% chlorine (bleach) or
water and mild detergent (dish soap) may be used. The cloth should be damp but not wet. Only the exterior of
the Tool may be cleaned. Do not attempt to clean the interior of the Tool as this will damage the Tool
’s
circuitry.
Do not use abrasive cleaners or cleaning products in aerosol cans as they will damage the Tool
’s
finish.
Tool Firmware Revision
Tool Battery Level
Antenna Resonant Frequency
Axis Identification
Capacitor Used