Ecleris EVERLUX MINI-PRO LED111 User Manual Download Page 8

USER MANUAL 

– EVERLUX - MINI PRO 

 

Page 

8

 of 

13

 

INMANUUIN169 Rev.2019.04.01 / ENG 

regulations in effect. 

If before connecting the EVERLUX any anomaly is detected, please contact the technical service 
department of your local distributor or the closest ECLERIS office, prior to its use.. 

5.2 OPERATION 

This section thoroughly describes the equipment start-up and functioning verification procedure. It 
is recommended to perform this procedure after installing the equipment and during maintenance 
inspections. 

This system must be used only for medical, paramedical or health professionals. 

 

1)  Connect the power supply cable to the equipment and the electrical network.  
2)  Turn the EVERLUX on by pressing the general on/off switch. 
3)  Turn the LED system on by turning the light intensity control (dimmer) clockwise. 
4)  Use this control to regulate the light intensity 

 

The EVERLUX light source has an over-temperature protection. LED system will turn off if 
it reaches a risky operation temperature. 

5)  Once  the  procedure  is  over,  turn  the  EVERLUX  off  in  order  to  maximize  the  LED  system 

lifetime and energy saving. 

6)  Should you use a video camera (Ecleris PROCAM), connect it to the video input connector 

(LED111 and LED121) 

7)  Connect  the  video  output  (rear  panel  BNC  or  S-Video  connector) to the video  input  of  the 

viewing device and turn the EVERLUX on. 

8)  Select the video input and video Standard in the viewing device. 

6.  MAINTENANCE 

In  order  to  keep  an  optimum  operation  of  the  EVERLUX  it  is  recommended  to  perform  the 
following procedures: 

  Periodic cleaning of all exterior components. 

  Connections inspection. 

  Annual technical inspection of the equipment.  

To  perform  any  revision  or  maintenance  work,  contact  your  local  distributor  or  the  closest 
ECLERIS office. 

6.1 CLEANING 

Before cleaning the EVERLUX, disconnect the power supply cable form the electrical network.   

EXTERNAL SURFACES 

Clean the external surface of the equipment using a piece of cotton or cloth moistened with mild 
detergent or soap. Use only adequate products for the external cleaning of electrical equipment, 
prepared in accordance with the manufacturer’s indications. 

Perform  the  cleaning  with  special  care,  avoiding  water  or  humidity  access  to  the 
equipment.  

Summary of Contents for EVERLUX MINI-PRO LED111

Page 1: ...hods without previous written authorization from ECLERIS Neither can it be reproduced multiplied or published through any electronic data processing form The information contained in this manual may be modified or broaden without previous notification without it representing any responsibility to ECLERIS ECLERIS USA 11801 NW 100th Rd Suite 17 Medley FL 33178 Phone 1 786 483 7490 CMC Medical Device...

Page 2: ... 5 3 3 1 MODEL LED110 5 3 3 2 MODEL LED111 5 3 3 3 MODEL LED120 6 3 3 4 MODEL LED121 6 3 3 5 REAR PANEL LED110 LED120 6 3 3 6 REAR PANEL LED111 LED121 6 4 INSTALLATION AND SET UP 7 5 REVISION AND OPERATION 7 5 1 REVISION 7 5 2 OPERATION 8 6 MAINTENANCE 8 6 1 CLEANING 8 6 2 ELECTRICAL CONNECTIONS 9 6 3 FUSE CHANGE 9 7 TECHNICAL SPECIFICATIONS 9 8 TROUBLESHOOTING 11 9 TECHNICAL SERVICE 12 10 WARRANT...

Page 3: ...cautions Please read the manual thoroughly before operating the equipment Should you have any doubt regarding the contents of the manual herein or the equipment operation and safety please contact your local distributor or the closest ECLERIS office 2 1 WARNINGS It is important to read this manual thoroughly before using the EVERLUX ECLERIS shall not be liable for the damages that may occur if the...

Page 4: ...horized personnel In case of any problem the system must be laid a plane surface and power cable disconnected The device must be used only by personnel informed of all this warnings Any use outside of the manufacturer specific intent use will exclude ECLERIS of all liability This user manual must be with the device for permanent review If the system is sold or donated the user manual must be deliv...

Page 5: ...uring systems etc 3 1 INTENDED USE The EVERLUX LED light source has been designed to be used with ECLERIS microscopes and colposcopes and in therapeutic and diagnostic endoscopy procedures 3 2 INTENDED USER The EVERLUX LED light source has been designed to be used by healthcare professionals 3 3 COMPONENTS EVERLUX light source includes the following components Main unit Power cord User manual It i...

Page 6: ...amera connector video input 3 3 5 REAR PANEL LED110 LED120 5 Earth connection 6 Power supply cable connector with fuse holder 7 Ventilation grid 3 3 6 REAR PANEL LED111 LED121 5 Earth connection 6 Power supply cable connector with fuse holder 7 Ventilation grid 8 S Video output SVHS 9 Composite Video output CVBS 10 Ecleris Endodigi Connector Image and video capture and other peripheral devices Sta...

Page 7: ...mperature The EVERLUX shall only be used in places with electrical installations that meet the local regulations in effect It shall always be connected to the grounding of the operating room or endoscopy tower through a ground cable The EVERLUX shall be connected to the voltage line through the power supply cable provided The power supply shall only be provided with a protected plug and the adequa...

Page 8: ... is over turn the EVERLUX off in order to maximize the LED system lifetime and energy saving 6 Should you use a video camera Ecleris PROCAM connect it to the video input connector LED111 and LED121 7 Connect the video output rear panel BNC or S Video connector to the video input of the viewing device and turn the EVERLUX on 8 Select the video input and video Standard in the viewing device 6 MAINTE...

Page 9: ...een the cable and the plug Oxide corrosion in connection pins 6 3 FUSE CHANGE To avoid loss of time by downtime of EVERLUX is recommended to have spare fuses 1 Disconnect the power supply cable from the electrical network and the equipment 2 Under the power supply cable connector there is a small rectangular plastic holder where the fuses are located Using a screwdriver and with a slight pressure ...

Page 10: ...d with a connectable power cable that assures proper electrical isolation of all terminals simultaneously from the main power MODEL LED110 LED111 LED120 LED121 FIBER OPTIC CONNECTOR Individual Interchangeable fiber optic cable adapter Built in revolving turret with multiple adapters VIDEO OUTPUT N A Composite Video CVBS 1 Vpp 75Ω BNC S Video Y C Jack mini DIN N A Composite Video CVBS 1 Vpp 75Ω BNC...

Page 11: ...Technical service Over temperature LED system protection activated Clean and clear the lateral ventilation grids Check if the fan is running Turn the equipment off immediately and contact the closest Technical Service Internal fuse burnt or defective board Contact the closest Technical service Light intensity is not the correct one Dimmer in minimum intensity Turn the dimmer clockwise to increase ...

Page 12: ...solved please contact your local distributor or the closest ECLERIS office and request technical service 9 TECHNICAL SERVICE The EVERLUX light sources must be repair and or maintained exclusively by qualified personnel trained by ECLEIRS Only the parts designed to be repair by ECLEIRS can be serviced ECLERIS recommends using the following procedures to speed up the repair service and replacement o...

Page 13: ...th the required Quality Control verification and has no manufacturing defects This guarantees the appropriate functioning under adequate use in agreement with the indications of this manual The warranty period is for one 1 year as from the date of purchase invoice ECLERIS shall not be liable for any damage caused to by the EVERLUX due to incorrect use negligence cleaning or incorrect storage Any r...