The  EVERLUX  shall  be  sent  in  its  original  packaging  so  that  it  is  not  damaged  during 
transportation. ECLERIS shall not be liable for damages that may occur during transportation, 
caused by deficient packaging. 

ECLERIS,  will  provide  if  required,  circuits,  drawings,  list  of  components,  descriptions  and 
instructions  for  calibration  or  other  useful  information  required  by  qualified  trained  technical 
personnel, to repair the parts of the device marked as reparable by the manufacture. 

10. 

WARRANTY 

ECLERIS states that the EVERLUX has complied with the required Quality Control verification 
and has no manufacturing defects. This guarantees the appropriate functioning under adequate 
use in agreement with the indications of this manual. The warranty period is for one (1) year as 
from the date of purchase invoice. 

ECLERIS shall not be liable for any damage caused to/by the EVERLUX due to incorrect use, 
negligence,  cleaning  or  incorrect  storage.  Any  repairs  made  by  personnel  not  authorized  by 
ECLERIS  will  void  this  warranty.  The  described  warranty  is  exclusively  valid  for  the  original 
buyer. 

The following parts are excluded from the warranty: 

  Damages resulting from connections to inadequate power supplies. 

Disregarding the warranty situation, all shipment from and to the authorized ECLERIS technical 
service facilities is the sole responsibility of the client. 

Summary of Contents for EVERLUX MINI-PRO LED111

Page 1: ...hods without previous written authorization from ECLERIS Neither can it be reproduced multiplied or published through any electronic data processing form The information contained in this manual may be modified or broaden without previous notification without it representing any responsibility to ECLERIS ECLERIS USA 11801 NW 100th Rd Suite 17 Medley FL 33178 Phone 1 786 483 7490 CMC Medical Device...

Page 2: ... 5 3 3 1 MODEL LED110 5 3 3 2 MODEL LED111 5 3 3 3 MODEL LED120 6 3 3 4 MODEL LED121 6 3 3 5 REAR PANEL LED110 LED120 6 3 3 6 REAR PANEL LED111 LED121 6 4 INSTALLATION AND SET UP 7 5 REVISION AND OPERATION 7 5 1 REVISION 7 5 2 OPERATION 8 6 MAINTENANCE 8 6 1 CLEANING 8 6 2 ELECTRICAL CONNECTIONS 9 6 3 FUSE CHANGE 9 7 TECHNICAL SPECIFICATIONS 9 8 TROUBLESHOOTING 11 9 TECHNICAL SERVICE 12 10 WARRANT...

Page 3: ...cautions Please read the manual thoroughly before operating the equipment Should you have any doubt regarding the contents of the manual herein or the equipment operation and safety please contact your local distributor or the closest ECLERIS office 2 1 WARNINGS It is important to read this manual thoroughly before using the EVERLUX ECLERIS shall not be liable for the damages that may occur if the...

Page 4: ...horized personnel In case of any problem the system must be laid a plane surface and power cable disconnected The device must be used only by personnel informed of all this warnings Any use outside of the manufacturer specific intent use will exclude ECLERIS of all liability This user manual must be with the device for permanent review If the system is sold or donated the user manual must be deliv...

Page 5: ...uring systems etc 3 1 INTENDED USE The EVERLUX LED light source has been designed to be used with ECLERIS microscopes and colposcopes and in therapeutic and diagnostic endoscopy procedures 3 2 INTENDED USER The EVERLUX LED light source has been designed to be used by healthcare professionals 3 3 COMPONENTS EVERLUX light source includes the following components Main unit Power cord User manual It i...

Page 6: ...amera connector video input 3 3 5 REAR PANEL LED110 LED120 5 Earth connection 6 Power supply cable connector with fuse holder 7 Ventilation grid 3 3 6 REAR PANEL LED111 LED121 5 Earth connection 6 Power supply cable connector with fuse holder 7 Ventilation grid 8 S Video output SVHS 9 Composite Video output CVBS 10 Ecleris Endodigi Connector Image and video capture and other peripheral devices Sta...

Page 7: ...mperature The EVERLUX shall only be used in places with electrical installations that meet the local regulations in effect It shall always be connected to the grounding of the operating room or endoscopy tower through a ground cable The EVERLUX shall be connected to the voltage line through the power supply cable provided The power supply shall only be provided with a protected plug and the adequa...

Page 8: ... is over turn the EVERLUX off in order to maximize the LED system lifetime and energy saving 6 Should you use a video camera Ecleris PROCAM connect it to the video input connector LED111 and LED121 7 Connect the video output rear panel BNC or S Video connector to the video input of the viewing device and turn the EVERLUX on 8 Select the video input and video Standard in the viewing device 6 MAINTE...

Page 9: ...een the cable and the plug Oxide corrosion in connection pins 6 3 FUSE CHANGE To avoid loss of time by downtime of EVERLUX is recommended to have spare fuses 1 Disconnect the power supply cable from the electrical network and the equipment 2 Under the power supply cable connector there is a small rectangular plastic holder where the fuses are located Using a screwdriver and with a slight pressure ...

Page 10: ...d with a connectable power cable that assures proper electrical isolation of all terminals simultaneously from the main power MODEL LED110 LED111 LED120 LED121 FIBER OPTIC CONNECTOR Individual Interchangeable fiber optic cable adapter Built in revolving turret with multiple adapters VIDEO OUTPUT N A Composite Video CVBS 1 Vpp 75Ω BNC S Video Y C Jack mini DIN N A Composite Video CVBS 1 Vpp 75Ω BNC...

Page 11: ...Technical service Over temperature LED system protection activated Clean and clear the lateral ventilation grids Check if the fan is running Turn the equipment off immediately and contact the closest Technical Service Internal fuse burnt or defective board Contact the closest Technical service Light intensity is not the correct one Dimmer in minimum intensity Turn the dimmer clockwise to increase ...

Page 12: ...solved please contact your local distributor or the closest ECLERIS office and request technical service 9 TECHNICAL SERVICE The EVERLUX light sources must be repair and or maintained exclusively by qualified personnel trained by ECLEIRS Only the parts designed to be repair by ECLEIRS can be serviced ECLERIS recommends using the following procedures to speed up the repair service and replacement o...

Page 13: ...th the required Quality Control verification and has no manufacturing defects This guarantees the appropriate functioning under adequate use in agreement with the indications of this manual The warranty period is for one 1 year as from the date of purchase invoice ECLERIS shall not be liable for any damage caused to by the EVERLUX due to incorrect use negligence cleaning or incorrect storage Any r...