49
scratch the printing head. Otherwise the printing head will be damaged. Load the recording paper
and close the cover when all traces of alcohol are evaporated.
4
)
Cleaning the Silicon Glue Alloy Shaft
Silicon glue alloy Shaft should be flat, smooth and free-of-dirt. Otherwise the ECG records will be
influenced. Wipe the shaft with damped soft cloth. Close the cover when it is air dried.
Prevent the liquid from leaking into the main unit while cleaning
Do not clean the unit and accessories with hard materials and avoid scratching the
electrodes.
6.2 Disinfection
It is recommended that disinfection should be only performed under the regulation of medical
situations or under other proper situations; disinfecting equipment must be disinfected before
cleaning the ECG machine. Do not use high-temperature, high-pressure steam, ionizing radiation
methods for disinfection.
Do not use chlorine-containing disinfectant for disinfection, such as chlorinated lime and
sodium hypochlorite etc.
6.3 Care and Maintenance
6.3.1 Battery Capacity, Charging and Replacement
Inappropriate operation may lead battery to be hot, ignited, exploded, damaged or capacity
fade. Before using the rechargeable lithium battery, read the operation manual carefully.
1
)
Charge
ECGMAC ECG is equipped with recharge control circuit together with built-in rechargeable
lithium battery. Because of the capacity consumption during storage and transport, the capacity of
battery is not full while using the first time. Battery-recharging should be suggested before first
usage. When connect with the mains supply, the battery will be recharged automatically. Then the
battery recharge indicator light and mains supply indicator light on. When the capacity of battery is
full, the battery recharge indicator light will be off.
2
)
Battery Capacity Indicator
When using the battery, there is a sign representing the battery capacity on the LCD screen. As
shown in the following figures,