reconnect them.
1
)
Electrodes fall off
①
When the lead wires are not connected to the patient or the instrument stably, ECG signals
can not be transmitted correctly, this message will be displayed for reminding;
②
When the polarization voltage exceeds the limit value, the message will be displayed for
reminding.
2
)
If the “Lead off” message is shown, handle it according to the following method:
①
Press “INST” button to eliminate the influence of the polarization voltage to amplifier. It
will get the amplifier back to normal quickly;
②
If pressing “INST” button does not eliminate the alarm, check the corresponding electrode
and its connection.
4.5 Battery Installation
See Figures below for installation:
1.
Turn the electrocardiograph upside-down, unscrew and remove the battery door as A enclosed;
3.
Tidy the leads, and put the battery into the battery pack horizontally (as Picture C)
Summary of Contents for EM-1203