ECD HCA1 Instruction Manual Download Page 12

12 

 

8.

 

Minimize exposure of meter and test reagents to heat above 32°C (90°F). 

9.

 

Dry the outside of the meter when testing is complete or before storage of the 
meter. 

10.

 

When running a DPD-1 Free Chlorine test AFTER a Total Chlorine DPD-4, or a 
HR Free Chlorine test, rinsing is very important to remove residual KI, which 
may interfere. 

11.

 

Each HCA1 Strip Micro is valid for ONLY one test. Discard strip after single use 
in regular refuse that is inaccessible to children and pets. 

12.

 

Each bottle of HCA1 Strip Micro contains, at minimum, the quantity of strips 
notated on the bottle. Due to the strip slitting process, you may find one or 
two extra strips that are noticeably smaller or larger in width than the normal 
strips in the bottle. These should be discarded. Using these strips may give 
incorrect results. 

13.

 

The HCA1 Chlorine Plus Meter is not compatible for use with DPD-1, DPD-3, 
and DPD-4 powder pillows, tablets, or liquids available from other 
manufacturers. Accurate results can only be guaranteed by using genuine 
HCA1 Micro strips or reagents (reorder information below). 

14.

 

Lab testing with the Chlorine Plus meter has shown that zeroing and 
measuring of the sample normally does not require any cell cover for accurate 
results, except in sunlight. To obtain optimal accuracy when testing with the 
meter outdoors (sunlight), use the Mixing Cap/Cell Cover when zeroing and 
reading the sample. 

15.

 

Remove batteries when meter is not used for more than a month (Warranty 
Requirement). 

HCA1 Chlorine Plus Meter Messages 

Message 

Description 

Corrective Action 

Hi 

In READ mode: test sample 
concentration is above the 
measurement range (test specific). 

Dilute and retest. Dilution Kit 
available 

Lo 

In READ mode: test sample 
concentration is below the 
measurement range (test specific). 

Sample value is below 
measurement range. 

Lo 

In ZERO mode: sample absorbance 
(due to a cloudy or colored sample 
or a dirty cell) is too high to zero, 
the meter will read “LO”. 

Dilute sample, filter sample, or 
clean cell. One of these options 
should remedy the problem. 
 

ER 

Excessive stray light detected. 
Normally this does not occur, even 
when testing in sunlight. 

Place the LIGHT BLOCKING CAP 
over the CELL for zeroing and for 
reading result. 

 

Low battery indication. 
 

Replace the batteries. 
 

 

Summary of Contents for HCA1

Page 1: ...HCA1 Chlorine Plus Photometer System Instruction Manual Ideal for the measurement of Free and Total Chlorine in Drinking Water Waste Water Environmental or other clear water samples...

Page 2: ...of test strips Ozone 1000040 7 Meter case and 2 bottles of test strips Peracetic Acid 9250500 Plastic case 1000570 HCA1 meter only 9260100 100 Test Strips Free Chlorine 0 01 11 0 ppm 9260101 100 Test...

Page 3: ...ll 6 Installing the Batteries 6 MENU CL1 DPD 1 Free Chlorine Test Procedure 7 MENU CL1 DPD 4 Total Chlorine Test Procedure 8 MENU O3 Ozone Test Procedure 8 MENU Cd4 Chlorine Dioxide Test Procedure 9 M...

Page 4: ...attery Life 2000 tests with alkaline batteries Electromagnetic Compliance Emitted Interference EN 61326 Immunity to Interference EN61326 Waterproof rating Exceeds IP67 Weight Instrument 140 g 5 oz Dim...

Page 5: ...result followed by 19 which is the 2nd latest result etc and finally 01 which is the oldest result retained Only the last 20 readings are stored in each menu This meter is able to store 160 results i...

Page 6: ...s compliant Likewise AWWA proportions are followed as required for Total Chlorine measurements using Potassium Iodide Components Free Chlorine AWWA 4500 CI G HCA1 DPD 1 Anhydrous DPD sulfate 1 5 1 5 A...

Page 7: ...press the MENU button until the display shows the parameter CL1 4 Rinse the CELL at least 3 times with the water sample you will be testing rinsing minimizes the potential for cross contamination from...

Page 8: ...mple Record result displayed this result is automatically stored in CL1 Menu 8 Press READ again and the meter will count down and display the next Total Chlorine result If this reading matches the pre...

Page 9: ...forth motion approx 2 strokes Sec Remove and discard the strip after 1 on the display disappears The cursor will move across the display while the meter prepares to measure the sample Ignore the resul...

Page 10: ...venient place and recap the test strip bottle 2 Press the ZERO ON button to power the meter on the display will show all annunciators then the current MENU selection followed by the last reading 3 Pre...

Page 11: ...temperature and Monochloramine concentration Ozone O3 Color similar to free chlorine reaction at all levels Peroxides Interference is possible pH Typical pH samples of potable water with a pH of 6 0...

Page 12: ...1 Micro strips or reagents reorder information below 14 Lab testing with the Chlorine Plus meter has shown that zeroing and measuring of the sample normally does not require any cell cover for accurat...