ECD HCA1 Instruction Manual Download Page 10

10 

 

4.

 

Rinse the CELL at least 3 times with the water sample you will be testing - 
rinsing minimizes the potential for cross contamination from a previous test. 
Finally, fill cell to capacity (4mL) with the water sample. 

5.

 

Press the ZERO/ON button. The cursor will move across the display followed 
by 0.00 PPM. Sample is ready for testing. 

6.

 

Dip the HCA1 Micro Strip PA (DPD-4) into the CELL and immediately press 
READ
. This starts the 20 SECOND countdown timer. During this time move the 
strip in a gentle back and forth motion (approx. 2 strokes/Sec). Remove and 
discard the strip after “1” on the display disappears. When testing outdoors 
(in sunlight), for best accuracy, use the Mixing Cap/Cell Cover when Zeroing 
and Reading the sample. 

7.

 

The cursor will move across the display while the meter prepares to measure 
the sample. Record result displayed (this result is automatically stored in PA5 
Menu).

 

After testing is completed, rinse cell immediately. 

MENU HR8 High Range Chlorine Test Procedure 

1.

 

Remove (1) Micro HR Strip from the bottle before beginning the test. Place it 
in a dry convenient place and recap the test strip bottle. 

2.

 

Press the ZERO/ON button to power the meter on; the display will show all 
annunciators, then the current MENU selection, followed by the last reading. 

3.

 

Press and re-press the MENU button until the display shows the parameter 
HR8. 

4.

 

Rinse the CELL at least 3 times with the water sample you will be testing - 
rinsing minimizes the potential for cross contamination from a previous test. 
Finally, fill cell to capacity (4mL) with the water sample. 

5.

 

Press the ZERO/ON button. The cursor will move across the display followed 
by 0.00 PPM. Sample is ready for testing. 

6.

 

Dip the HCA1 Micro Strip HR into the CELL and immediately press READ. This 
starts the 20 SECOND countdown timer. During this time move the strip in a 
gentle back and forth motion (approx. 2 strokes/Sec). Remove and discard the 
strip after “1” on the display disappears. When testing outdoors (in sunlight), 
for best accuracy, use the Mixing Cap/Cell Cover when Zeroing and Reading 
the sample. 

7.

 

The meter will begin counting up for 120 seconds, at the end of which the 
cursor will move across the display while the meter prepares to measure the 
sample. Record result displayed (this result is automatically stored in HR8 
Menu).

 

After testing is completed, rinse cell immediately. 

 

 

 

 

Summary of Contents for HCA1

Page 1: ...HCA1 Chlorine Plus Photometer System Instruction Manual Ideal for the measurement of Free and Total Chlorine in Drinking Water Waste Water Environmental or other clear water samples...

Page 2: ...of test strips Ozone 1000040 7 Meter case and 2 bottles of test strips Peracetic Acid 9250500 Plastic case 1000570 HCA1 meter only 9260100 100 Test Strips Free Chlorine 0 01 11 0 ppm 9260101 100 Test...

Page 3: ...ll 6 Installing the Batteries 6 MENU CL1 DPD 1 Free Chlorine Test Procedure 7 MENU CL1 DPD 4 Total Chlorine Test Procedure 8 MENU O3 Ozone Test Procedure 8 MENU Cd4 Chlorine Dioxide Test Procedure 9 M...

Page 4: ...attery Life 2000 tests with alkaline batteries Electromagnetic Compliance Emitted Interference EN 61326 Immunity to Interference EN61326 Waterproof rating Exceeds IP67 Weight Instrument 140 g 5 oz Dim...

Page 5: ...result followed by 19 which is the 2nd latest result etc and finally 01 which is the oldest result retained Only the last 20 readings are stored in each menu This meter is able to store 160 results i...

Page 6: ...s compliant Likewise AWWA proportions are followed as required for Total Chlorine measurements using Potassium Iodide Components Free Chlorine AWWA 4500 CI G HCA1 DPD 1 Anhydrous DPD sulfate 1 5 1 5 A...

Page 7: ...press the MENU button until the display shows the parameter CL1 4 Rinse the CELL at least 3 times with the water sample you will be testing rinsing minimizes the potential for cross contamination from...

Page 8: ...mple Record result displayed this result is automatically stored in CL1 Menu 8 Press READ again and the meter will count down and display the next Total Chlorine result If this reading matches the pre...

Page 9: ...forth motion approx 2 strokes Sec Remove and discard the strip after 1 on the display disappears The cursor will move across the display while the meter prepares to measure the sample Ignore the resul...

Page 10: ...venient place and recap the test strip bottle 2 Press the ZERO ON button to power the meter on the display will show all annunciators then the current MENU selection followed by the last reading 3 Pre...

Page 11: ...temperature and Monochloramine concentration Ozone O3 Color similar to free chlorine reaction at all levels Peroxides Interference is possible pH Typical pH samples of potable water with a pH of 6 0...

Page 12: ...1 Micro strips or reagents reorder information below 14 Lab testing with the Chlorine Plus meter has shown that zeroing and measuring of the sample normally does not require any cell cover for accurat...

Reviews: