FCA
‐
22
Instruction
Manual
Page 8
Start
Up
Supply
sample
flow
to
the
system.
Adjust
the
sample
flow
until
the
water
is
overflowing
into
the
central
overflow
tube.
At
higher
flow
rates
the
excess
water
will
overflow
down
the
drain.
A
minimum
flow
of
10
gal/min
is
required
to
maintain
the
overflow
condition.
Verify
the
sample
flow
is
passing
through
both
flow
cells
and
returning
to
the
drain.
The
Chlorine
Sensor
requires
1
‐
1
½
hours
of
polarization
by
the
instrument
before
accurate
measurements
can
be
made.
Allow
the
system
to
run
at
least
1
hour
before
proceeding
to
the
Chlorine
Sensor
calibration
below.
Calibration
Temperature
Calibration
The
Temperature
sensor
inside
the
PHS10
sensor
was
calibrated
at
the
factory
before
shipping.
Verify
the
displayed
temperature
agrees
with
the
actual
temperature.
If
the
value
deviates
by
more
than
2°
C
then
adjust
the
Temperature
Trim
in
the
Configuration
&
Trim
Menu
to
the
correct
value.
For
a
detailed
description
see
Temperature
Calibration
in
the
C
‐
22
manual.
Temperature
Calibration
Screen
Shots
#1
Enter
the
Configure/
Trim
Menu
Scroll
to
the
Configure
Menu
using
the
MENU
SELECT
key
and
enter
the
Menu
by
simultaneously
pressing
both
Horizontal
CALIBRATE
keys.
#2
Scroll
through
the
Passwords
Screen
Press
the
MENU
SELECT
down
Key
four
times
to
the
next
screen.
#3
Scroll
through
the
Relay
Assign
Screen
Press
the
MENU
SELECT
down
Key
four
times
to
the
next
screen.
#4
Choose
°C/°F
&
temp
cal
Select
the
Menu
by
Simultaneously
pressing
both
Horizontal
CALIBRATE
keys.
#5
Choose
the
Trim
°C
Menu
Select
the
Menu
by
Simultaneously
pressing
both
Horizontal
CALIBRATE
keys.
#6
Adjust
the
Trim
to
calibrate
the
reading
Using
the
vertical
CALIBRATE
keys
adjust
the
trim
to
change
the
Displayed
temperature