Safety information
1.1. Dangers of maloperation
The
WSPS 05
underwent a safety and acceptance test based on IEC 1010-1/EN 61010
– 1
part 1. Integrated electronic and hydraulic elements serve protection purposes and ensure
safe handling if used the conventional way. False operation or abuse, as well as
insensitivity for the device’s limits and safety instructions will result in immediate danger for
the following:
the operator’s health and life,
the
WSPS 05
and all systems it is connected to,
the accuracy of the
WSPS 05
,
the environment.
This manual includes information and safety instructions, which guarantee safe handling
and support your devices unobjectionable condition.
Therefore it is a necessity for all people who operate, install or maintain this device to mind
this manual.
1.2. Conventional use
With this
WSPS 05
you have purchased an effective diagnostic system for the saturation
level of water in oil. It operates very reliable and meets the requirements of daily use.
The set is recommended and tested for the use with:
Standard hydraulic oils:
HL, HLP
Triglyceride:
HETG
Aircraft hydraulics:
Mil H 5606
Synthetic ester:
HEES
Lubrication oils:
CLP
Limits:
Maximum acceptable pressure at the sensor
25 bar
(362,5 psi)
Maximum oil temperature
90° C
(194° F)
Maximum acceptable flow rate at the sensor
2 m/s
(6,5 ft/s)
The
WSPS 05
should always be operated with 12...30 V DC!