INM MTL 130-0200 Rev 4

15

Table 1

 - Common groups (continued)

Send

Function

Reply (verbose)

Reply (terse)

Limits (where applicable)

Notes

AxPy

Parameters.

P9 Terse=0

P9=1

0 or 1

1 to set instrument into terse 
mode – 0 to set verbose

P8 A2 Mode=1

P8 =1

0 to 3

0, 1, 2, 3 in terse - Off, High, 
Low, status in verbose

P7 A2 Hyst=1.0%

P7 =1.0

1 to 10 (1dp)

Alarm hysteresis

P6 A2 Level=5.0%

P6 =5.0

0 to 100

Alarm trip point

P5 A1 Mode=1

P5 =1

As alarm 2

As alarm 2

P4 A1 Hyst=1.0%

P4 =1.0

As alarm 2

As alarm 2

P3 A1 Level=5.0%

P3 =5.0

As alarm 2

As alarm 2

P2 4mA=0%

P2 =0

Will show 0V or 4mA 
depending on variant.

P1 20mA=50%

P1 =50

0 to 100

Zero and FSD limits are 
interactive.

AxRy

Reading  

* see (8) below

R5 Comp2=N/A

R5 =0

R4 Temp=Normal

R4 =1

Heater state – see also heater 
errors.

R3 Alarm2=Normal

R3 =0

0 to 2

Alarm state – Off, Normal, 
ALARM or N/A (if not fitted)

R2 Alarm1=Normal

R2 =0

0 to 2

In terse mode alarm state is 
1 in alarm and 0 for all other 
states

R1 Conc=5.00%

R1 =5.00

Instrument range and resolution

Measure and Concentration

Notes on readings:- In over-range condition (110% of instrument span) xxxxx will be ‘+++++’ (e.g.  ‘R1 Conc= +++++’). Similarly  in under-range 
condition (-5% of instrument span) xxxxx will be ‘-----’. In the case of a system error R1 will be replaced by the error code e.g. ? 72 In verbose 
mode a brief fault description will be appended. ‘Conc’ & ‘%’ are determined by sensor type and unit – see ‘U’ below. Resolution is applica-
tion dependant. Where a fault is present in the secondary reading it will be indicated by a text description in verbose mode and by a code in 
the case of Zr units (heater does not return a numeric value in R4) or by -999 or +999 to indicate downscale or upscale failures respectively.

AxUy

Unit 
(read only)

U13 Test  Flags=0

U13=0

U12 Factory Flags=0

U12=0

U11 Output=mA*

U11=1

Analogue output: 0=4/20mA, 
1= 0/1V,  2 = 0/5V, 3=0/20mA

U10 Sens 2 Ch=1

U10=1

Sens 2 ADC chan (9)

U9 R2 unit=mV

U9=2

U8 R2 type=T/C

U8 =14

U7 R1 Ch=1

U7 =1

U6 R1 unit=%

U6 =1

U5 R1 type=Z

U5 =13

U4 F/w rev=0.24

U4 =0.24

U3 F/w p/n=290-6031

U3 =290-6031

U2 S/n = I-700123

U2=I-700123

U1 Addr=0

 U1 =0

Sensor Card Address

Summary of Contents for Crouse-Hinds Z130

Page 1: ...Z130 MTL zirconia oxygen analyser February 2017 INM MTL 130 0200 Rev 4 Instruction manual MTL gas analysers systems...

Page 2: ...LARATION OF CONFORMITY A printed version of the Declaration of Conformity has been provided separately within the original shipment of goods However you can find a copy of the latest version at http w...

Page 3: ...3 2 18 Enclosure details 3 3 INSTALLATION 4 3 1 Unpacking and visual checking 4 3 2 Calibration overview 4 3 3 Sample 4 3 3 1 Key requirements 4 3 3 2 General advice Sample 5 3 3 3 Measuring low leve...

Page 4: ...iv INM MTL 130 0200 Rev 4 THIS PAGE IS LEFT INTENTIONALLY BLANK...

Page 5: ...labelling Normal relates to process normal and not the electrical rest position of the relays In process normal the relays are energised 1 2 Manual symbols The following methods are used in this manua...

Page 6: ...pm 0ppm to 9 9ppm 0 1ppm 2 5 Stability Better than 2 of reading or 0 5ppm month whichever is greater 2 6 Sample flow Between 100 and 500 ml min for optimum operation 2 7 Speed of response T 90 less th...

Page 7: ...programmable for Mode HIGH LOW or OFF Level full range of instrument Hysteresis 0 to 10 of set point Volt free C O contacts rated at 48v 0 5A AC or DC normally energised 2 14 Serial Communication Por...

Page 8: ...FLAMMABLE MIXTURES ARE ALLOWED TO COME INTO CONTACT WITH THE MEASURING CELL AS THIS MAY CAUSE IGNITION OF THE GAS GASES CONTAINING HALOGENS SULPHUR OR SILICON MUST BE AVOIDED IT IS THE RESPONSIBILITY...

Page 9: ...und joints and couplings When measuring concentrations of less than 500ppm oxygen very small leakages can noticeably affect the readings for ppm concentrations the piping must be of hard plastic or me...

Page 10: ...ent 2 2 3 TX Data transmitted from instrument 3 3 2 3 4 3 Alarm and Analogue Output connections See figure 2 CAUTION The recommended cable used for external connection is double insulated WARNING Alth...

Page 11: ...the same principle use the or arrows to scroll through the parameters By pressing Edit and the x in the xP nnnn will flash then using the or arrows to edit the value Pressing Edit saves change while E...

Page 12: ...8 INM MTL 130 0200 Rev 4 Figure 3 130 series menu...

Page 13: ...imately the same oxygen concentration as the normal sample As with most instruments of this type it is important to have a reasonable concentration difference between the two calibration points For th...

Page 14: ...rect level then press the Edit button to store the calibration setting or Cal to cancel To return the unit to measuring mode press and release the Cal button otherwise the or arrows until the display...

Page 15: ...o be adequate WARNING Opening the analyser case will expose you to mains voltage Disconnect the analyser from the mains supply and any potentially hazardous process wiring before opening the case 6 1...

Page 16: ...d by the Nernst equation Cell output 2 303RT log P1 4F P2 where R molar gas constant T absolute temperature of cell in K F Faraday constant P1 partial pressure of oxygen in the reference air in most c...

Page 17: ...age terminator CR LF 9 A response message maximum duration is defined as 3 second This is the maximum time from the first valid response character to the last message terminator CR LF This is to allow...

Page 18: ...w counts D4 ADC 1 12345cts D4 12345 ADC raw counts D3 Sens 3 N A D3 0 N A D2 Sens 2 11 56mV D2 11 56 Thermocouple mV D1 Sens 1 11 55mV D1 11 55 Oxygen cell mV AxEy Error logs read only except E9 E9 Cl...

Page 19: ...Conc 5 00 R1 5 00 Instrument range and resolution Measure and Concentration Notes on readings In over range condition 110 of instrument span xxxxx will be e g R1 Conc Similarly in under range conditio...

Page 20: ...legal for this board 71 79 CRC error NVRAM errors Error 71 user parameters CRC error if this area is restored user calibration etc is LOST Error 71 will be reported in response to ANY read request unt...

Page 21: ...INM MTL 130 0200 Rev 4 17 THIS PAGE IS LEFT INTENTIONALLY BLANK...

Page 22: ...18 INM MTL 130 0200 Rev 4 THIS PAGE IS LEFT INTENTIONALLY BLANK...

Page 23: ...INM MTL 130 0200 Rev 4 19 THIS PAGE IS LEFT INTENTIONALLY BLANK...

Page 24: ...rozavodskaya Str 33 Building 4 Moscow 107076 Russia Tel 7 495 981 3770 Fax 7 495 981 3771 E mail mtlrussia eaton com SINGAPORE Cooper Crouse Hinds Pte Ltd No 2 Serangoon North Avenue 5 06 01 Fu Yu Bui...

Reviews: