Calibration and settings (by the Easytech Technical Service)
Calibration consists of a procedure that allows to evaluate the error of the machine on the
measure of a variable whose value is significant for the purposes of treatment, in order to
determine whether to accept it (as negligible) or to correct it through calibration or to repair
if it is a failure. An error may in fact be generated as a result of drift over time or as a result
of a malfunction.
The setting consists instead in a series of actions to ensure that the machine measure
correctly again when the error is not caused by the presence of a fault, but only by a drift or
by the replacement of a component. In the event that it is instead a failure, first will be
performed the repairing (replacement of a component), then the calibration and the setting
if necessary.
Repairs and settings are carried out by
Easytec
h
Technical Support.
7.4
Fuses replacement
Fuses are located in the group of the power supply; the operations for replacing a fuse are:
1)
Rotate the fuse holder with the aid of a tool and remove it;
2)
Replace the fuse;
3)
Put the fuse holder with new fuse into the hole and, by using a tool, rotate it until it
locks.