Easytech s.r.l. – User Manual T CaRe Compact
rev. D
Pag. 22 di 36
4.3
Treatment
This section describes the actions done from user interface via touchscreen. For more
information about the treatment (electrode placement and execution of the procedure),
refer to the followings chapters.
The screen appearing after the logo is displayed allows to manually select the parameters
for the treatment and, after selection, activate treatment.
From this window, with the
M
key, you can access the
set-up
(section 4.5).
4.4
Manual Settings
From the menu you select
(see Figure 4-1 and Figure 4-2):
resistive or capacitive mode of operations;
the energy output mode. You can have
continuous mode
(symbol
) or
pulsed mode
(symbol
);
the working frequency (500 kHz)
the treatment duration (from 10 sec. to 60 minutes) or the energy set (1 to 999 KJ);
You set one of the two parameters, after selecting the mode of operation,
time
or
energy
;
the frequency for the
pulsed
mode. The selection is done using the arrows; you can
set a value from 1 to 5 Hz, step 1 Hz. The selected value is displayed next to the
arrows;
The
power to be delivered, using the arrows on the left side of the display; on the
upper left corner you can see power value in percentage.
Figure 4-1- Manual setting of operating parameters – Time mode