43
The "Ht" and thermometer
icon appears. Temperature is
too high, outside the required
range of 10°C ~40°C (50°F
~104°F). This alerts users that
an incorrect result may occur
if the test continues.
Action:
Relocate the meter to a
location with temperature
between 10°C ~40°C (50°F
~104°F).
The "Lt" and thermometer
icon appears. Temperature is
too low, outside the required
range of 10°C ~40°C (50°F
~104°F). This alerts users that
an incorrect result may occur
if the test continues.
Action:
Relocate the meter to a
location with temperature
between 10°C ~40°C (50°F
~104°F).
System fault
Action:
Replace the
batteries first. If 005
appears again, please
contact your local dealer.