47
9. Critically ill patients should not be tested with home-use blood glucose meters.
10. DO NOT use serum or plasma sample for
EasyMax
®
R2N
SMBG System.
11. DO NOT use anticoagulant NaF or potassium oxalate for venous sample preparation.
12. Interfering Substances depend on the concentration. The below substances up to the test
concentration will not affect the test results.
4.4 mmol/L
(80 mg/dL)
13.8 mmolL
(250 mg/dL)
27.7 mmol/L
(500 mg/dL)
Ascorbic Acid, 0.227 mmol/L (4 mg/dL) 10.89%
-1.76%
4.55%
Ibuprofen, 2.424 mmol/L (50 mg/dL) 3.10%
2.88%
4.62%
L-Dopa, 0.091 mmol/L (1.8 mg/dL) 10.59%
7.91%
4.90%
Sodium Salicylate, 3.123 mmol/L (50 mg/dL)
-2.59%
9.42%
-0.84%
Tetracycline, 0.034 mmol/L (1.5 mg/dL) -5.32%
3.81% 3.20%
Tolbutamide, 3.699 mmol/L (100 mg/dL) -2.60%
12.30% 0.89%
Bilirubin-unconjugated, 0.041 mmol/L
(2.4 mg/dL)
-2.52% 4.05% -0.23%
Uric acid, 0.476 mmol/L (8 mg/dL) 2.71%
9.55%
-1.75%
Xylose, 0.266 mmol/L (4 mg/dL)
7.56%
0.76%
7.49%
Glucose
Level
Bias
Concentrations
of the interferant tested