background image

10  Setting the Sound On/OFF

Step 5 

On pressing the 

 or 

 button, the screen will display ‘On’ or ‘OFF’. Press the 

S

 

button to confirm the selection.

The meter will beep in the following instances if set to On.

•  When you push a button to turn on the meter,

•  When the test strip is inserted in the meter,

•  When the blood sample is absorbed into the test strip and 

the test starts,

•  When the test result is displayed,

•  When you press and hold the 

 button to set the post-

meal (PP2) alert,

•  When it is time for a pre-set blood glucose test.
If the sound is set to OFF, none of the sound functions will 

work.

After setting the sound, press the 

S

 button to  

go to the next step.

 

 Note

The 

 symbol is displayed only when the sound is set to OFF.

11  Turning on the Strip Expiration Date Indicator

Step 6 

This setting allows you to turn the strip expiration date indicator on or off. This 

setting turns the function on or off only. See page 14 to set the strip expiration 

date. When ‘EP’ appears on the screen, press the 

 or 

 button. The screen will 

display ‘On’ or ‘OFF’. Press the 

S

 button to confirm the setting. If you do not want 

to set the indicator, press the 

S

 button while the screen displays ‘OFF’.

User Manual

INDEX

01  Important Information: Read This First!

For optimum safety and benefits, please read the entire manual contents before 

using the system. 

Intended use:

CareSens S Fit BT Blood Glucose Monitoring System is intended for the 

quantitative measurement of glucose in fresh capillary whole blood from the 

fingertip.

The CareSens S Fit BT Blood Glucose Monitoring System is intended for use 

outside the body (

in vitro

 diagnostic use) and is intended for use as an aid 

to monitor the effectiveness of diabetes management. The system is for 

self-testing or healthcare professional use and should not be used for the 

diagnosis of or screening for diabetes or for testing newborns. 

PGUAA-0000380 REV0 2023-03

•  An abnormally high or low red blood cell count (hematocrit level over 65 % or 

below 15 %) may produce inaccurate results.

•  If your test result is below 3.3 mmol/L or above 13.3 mmol/L, consult a 

healthcare professional immediately.

•  Inaccurate results may occur in severely hypotensive individuals or patients 

in shock. Inaccurate low results may occur for individuals experiencing 

a hyperglycemic hyperosmolar state, with or without ketosis. Critically ill 

patients should not be tested with blood glucose meters.

•  Inaccurate results may occur in patients undergoing oxygen therapy.

If you need assistance, please contact the Spirit Healthcare 24/7 Freephone 

Support Line at 0800 881 5423 or go to www.spirit-healthcare.co.uk.

02 Specifications

Product specifications

Measurement range

1.1–33.3 mmol/L

Sample size 

Minimum 0.5 μL

Test time

6 seconds

Sample type

Fresh capillary whole blood

Calibration

Plasma-equivalent

Assay method

Electrochemical

Battery life

1,000 tests

Power

One 3.0 V lithium battery  

(disposable, type CR2032) 

Memory

1,000 test results

Size

95 x 49 x 17.5 mm

Weight

51.5 g (with battery)

Bluetooth

®

 technology

•  Frequency range: 2.4–2.4835 GHz

•  Operating range distance: 

maximum 10 meters (unobstructed)

•  Operating channels: 40 channels

•  Security encryption: 128-bit AES 

(Advanced Encryption Standard)

Operating ranges

Temperature

5–45 °C

Relative humidity

10–90 %

Hematocrit

15–65 %

Storage/Transport Conditions 

Temperature

Meter (with battery)

0–50 °C

Test strip

1–30 °C

Control solution

8–30 °C

Relative Humidity

Test strip 

10–90 %

03  CareSens S Fit BT Blood Glucose Monitoring 

System

CareSens S Fit BT Blood Glucose Monitoring System includes the following 

items:

* CareSens S Fit BT Blood Glucose Meter           * Instructions for Use

CareSens S Fit BT Blood Glucose Monitoring System may include the 

following items:

* CareSens S Blood Glucose Test Strips

* Lancing Device

* Battery

* Logbook

* Lancets

* Carrying Case

•  Check all the components after opening the CareSens S Fit BT Blood Glucose 

Monitoring System package. The exact contents are listed on the main box.

•  Please contact the Spirit Healthcare 24/7 Freephone Support Line for a free data 

transmission cable at 0800 881 5423 or go to 

www.spirit-healthcare.co.uk

.

04 Inserting or Replacing the Battery

The CareSens S Fit BT meter uses one 3.0 V lithium battery. Before using the 

meter, check the battery compartment and insert a battery if empty.

When the 

 symbol appears on the display while the meter is in use, the 

battery should be replaced as soon as possible. The test results may not be saved 

if the battery runs out.

Step 1 

Make sure the meter is turned off. Push the 

cover in the direction of the arrow to open the 

battery compartment.

Step 2 

Remove the used battery. Slip your thumb 

finger and pulling it out with your index and 

thumb finger as shown in the diagram. Insert 

a new battery with the + side facing up and 

make sure the battery is inserted firmly.

Step 3 

Place the cover on the battery compartment. 

Push it down until you hear the tab click into 

place.

 Note

Removing the meter battery will not affect your stored results. However, you 

may need to reset your meter settings.  See page 10.

05 Caring for Your System

Use a soft cloth or tissue to wipe the meter exterior. If necessary, dip the soft 

cloth or tissue in a small amount of alcohol.

Do not use organic solvents such as benzene or acetone, or household and 

industrial cleaners that may cause irreparable damage to the meter.

Caution:

•  Do not expose the meter to direct sunlight, heat, or excessive humidity for an 

extended period of time. It is recommended to store and use the test system 

indoors.

•  Do not let dirt, dust, blood, or water enter into the meter’s test strip port.

•  Do not drop the meter or submit it to strong shock.

•  Do not try to fix or alter the meter in any way.

•  Keep the meter away from strong electromagnetic field sources, such as cell 

phones and microwave ovens.

•  Store all the meter components in the carrying case to prevent loss and help 

keep the meter clean.

•  Avoid getting any liquid or moisture in the test strip vial. This can affect the 

test strips and cause inaccurate test results. 

•  Do not apply samples other than capillary whole blood or control solution to 

the test strip.

•  Store the meter in a cool and dry place between 0–50 °C.

Disposal of the meter

If you need to throw your meter away, you should follow existing policies and 

procedures of your own country or region. For information about correct 

disposal, please contact your local council or authority. If you need assistance, 

contact the Spirit Healthcare 24/7 Freephone Support Line at 0800 881 5423 or 

go to 

www.spirit-healthcare.co.uk.

06 CareSens S Blood Glucose Test Strip

The CareSens S Fit BT Blood Glucose Monitoring System measures blood 

glucose quickly and accurately. It automatically absorbs the small blood sample 

applied to the tip of the strip.

Contact bars

Gently push the test strip, with its contact bars 

facing up, into the test strip port of meter

Tip to apply blood sample

Apply blood sample here for testing

Confirmation window 

Check here to see whether sufficient blood 

sample has been applied

Warning!

•  The CareSens S Test Strips should be used only with fresh capillary whole 

blood samples.

•  Do not reuse test strips.

•  Do not use test strips past the expiration date.

•  Test strips in new, unopened vials and test strips in vials that have been 

opened can be used up until the expiration date printed on the test strip box 

and vial label if the test strips are used and stored according to its storage and 

handling methods.

•  Store test strips in a cool and dry place at a temperature between 1–30 ˚C and 

10–90 % relative humidity. 

•  Keep test strips away from direct sunlight or heat and do not freeze.

•  Store test strips only in their original vial.

•  Avoid getting any liquid or moisture in the test strip vial. This can affect the 

test strips and cause inaccurate test results.

•  Do not apply samples other than capillary whole blood or control solution to 

the test strip.

•  Close the vial tightly after taking out a test strip for testing and use the strip 

immediately.

•  Handle test strips only with clean and dry hands.

•  Do not bend, cut, or alter test strips in any way.

•  For detailed storage and usage information, refer to the CareSens S test strip 

package insert.

 Caution

•  Keep the meter and testing supplies away from young children.

•  Drying agents in the vial cap may be harmful if inhaled or swallowed and 

may cause skin or eye irritation.

07 CareSens S Fit BT Blood Glucose Meter

Data Port

Used to transfer 

data from the 

meter to a 

computer with a 

cable

Display 

Shows results and 

messages

S Button

Turns the meter 

on/off, confirms 

menu selections, 

and changes 

information

,

 Button

Turns the meter on, 

selects or changes 

information

Test Strip Port

Insert test strip here

 Note

•  The cable for data management software can be ordered separately. Please 

contact your authorised i-SENS sales representative.

•  The unit of measurement is fixed and it cannot be changed by the user.

08  CareSens S Fit BT Blood Glucose Meter Display

1

13

18

17

8

2 3 4

19

5 6 7

9

10

11

12

14

15 16

1

 

Test results:

 test results displaying panel

2

  

Memory symbol:

 appears when test results stored in 

the memory are displayed

3

  

Setting symbol:

 appears when in SET mode

4

 

 PP2  alarm:

 appears when the post-meal alarm has 

been set

5

  

Bluetooth Symbol

6

  

Mute symbol:

 appears only when the sound is set 

to OFF

7

  

Control Solution flag:

 appears when the control solution test results are 

saved or displayed

8

  

KET symbol:

 appears when the test result is greater than 13.3 mmol/L

9

  

Hi:

 appears when the test result is greater than the selected hyperglycemia 

level

10

  

Smile symbol:

 appears when the test result is within the selected normal 

blood glucose range

11

  

Lo:

 appears when the test result is lower than the selected hypoglycemia 

level

12

  

alarm:

 appears when the time alarm has been set

13

  

mmol/L, mg/dL:

 unit for measuring blood glucose

14

  

Pre-meal test flag:

 used for tests done before eating

15

  

Post-meal test flag:

 used for tests done after eating

16

  

Fasting test flag:

 used for tests done after fasting for at least 8 hours

17

  

Battery symbol: 

indicates meter battery is running low and needs to be 

replaced

18

  

Blood insertion symbol:

 indicates meter is ready for the application of a 

drop of blood or control solution

19

  

Month/Day/Hour/Minute:

 appears date and time

 Note

It is recommended to check if the display screen on the meter matches the 

illustration above every time the meter turns on. Do not use the meter if the 

display screen does not exactly match the illustration as the meter may show 

incorrect results. 

09 Setting Up Your System

Press and hold the 

S

 button for 3 seconds to enter SET mode. After all settings 

are finished, press and hold the 

S

 button for  

3 seconds to turn off the meter. Press the 

 or 

 button to change values. Press 

and hold the 

 button to scroll faster.

Step 1 

Entering the SET Mode

Press and hold the 

S

 button for  

3 seconds to enter SET mode. After all 

the segments flash across the screen, 

‘SET’ will show up.

Press the 

 or 

 button to select ‘YES’ 

and press the 

S

 button to go to the next 

step.

If you do not want to connect your meter to your smartphone, press the 

S

 button 

when the screen display ‘bT’ while ‘OFF’ is blinking on the botton of the screen. 

Then, the meter will go to Step 4 Adjusting the Date and Time mode.

 Note

Follow steps 2–3 to pair your meter and smartphone. Pairing allows the meter 

to communicate wirelessly with your smartphone. Ensure that devices are 

within the maximum Bluetooth range (10 metres). Before pairing your meter 

and smartphone, download and install the SmartLog mobile app on your 

smartphone.

Setting Up Bluetooth
Step 2 

Bluetooth Pairing

  

Smartphone:

 If you want to pair (connect) your meter to your smartphone, 

launch the SmartLog mobile app and find the Accessories menu on your 

smartphone.

  

Smartphone:

 Select the CareSens S Fit model from the meters list, and then 

tap 

Bluetooth Register

 > 

Next

.

  

Meter:

 With your meter turned off, press and hold the 

 button for three seconds to enter pairing mode. The 

Bluetooth symbol and ‘bT’ will appear and ‘YES’ will blink 

at the bottom of the screen. Press the 

S

 button to select 

‘YES’.

  

Smartphone/Meter:

 Tap 

1

  on your smartphone to set the 

Bluetooth pairing type setting while Bluetooth symbol 

and ‘tYPE 1’ are displayed on the meter’s screen.

Blood Glucose Meter

xxxxxxxxxxxx

Serial 

number

  

Smartphone/Meter:

 Follow the on-screen 

instructions on SmartLog mobile app to begin 

searching your meter. Look for ‘CareSens’ and 

the last 4 characters of the meter serial number 

on the SmartLog mobile app screen to correctly 

identify your meter. Touch your meter’s ID 

(CareSens XXXX) on the SmartLog mobile app 

screen. 

 Note

Some content or menus may differ depending on your smartphone’s 

operating system or SmartLog version.  

Step 3 

Entering the PIN number

  

Meter/Smartphone :

 The   symbol and 6-digit PIN 

number will appear on the meter’s screen.

  

Smartphone:

 Enter the PIN number into the SmartLog mobile app and touch 

‘OK’. Make sure the PIN you enter on your smartphone matches the PIN on 

your meter screen.

  

Meter:

 When your meter and smartphone are paired and 

connected, the meter will display ‘SYNC’ and the saved 

test results will be transferred to your smartphone.

  

Meter:

 When the data transfer is finished, the meter will 

be display ‘End’ on the screen and it will be automatically 

turned off after 3 seconds. 

If the data transfer is failed, the meter will display ‘FAIL’ 

and it will be automatically turned off after  

3 seconds. Then, repeat Step 2 to 5 to connect your meter 

with smartphone.

 Note

Some smartphones, especially those that are not tested or approved by 

i-SENS, may be incompatible with your meter. Visit 

www.i-sens.com/smartlog

 

for more information about supported smartphones. You can also scan the QR 

code on the back cover of this user manual.

Step 4 

Adjusting the Date and Time

Press the 

 or 

 button to adjust the date and time. Following images will 

appear in order of year, month, date, time format, hour, and minutes. Press the  

S

 button to confirm your selection and to go to the next step.

 Note

If the pre-set expiration date expires, 

the meter will display ‘EP’ when 

the test strip is inserted. ‘EP’ shows 

alternately also when the test result 

is displayed right after the test.  

If the expiration date is set to 

October of 2025, the meter will 

display ‘EP’ at the beginning of 

November, 2025.

12  Setting the Hypoglycemia (Lo) Indicator

Step 7 

This setting allows you to select the desired level for the 

hypoglycemia indicator (possible low blood sugar). You will 

be alerted any time your test result is lower than the selected 

level. Press the 

 or 

 button until the desired hypoglycemia 

level between 1.1 and 5.0 mmol/L appears. Then, press the 

S

 

button to confirm the level and to go to the next step.

 Note

If the test result is lower than the pre-set 

hypoglycemia level, the meter will display ‘Lo’

 Caution

Ask your healthcare professional to help you decide what your hypoglycemia 

level is before setting your level.

13  Setting the Hyperglycemia (Hi) Indicator

Step 8 

This setting allows you to select the desired level for the 

hyperglycemia indicator (possible high blood sugar). You 

will be alerted any time your test result is higher than the 

selected level. Press the 

 or 

 button until the desired 

hyperglycemia level between 6.7–19.4 mmol/L appears. 

Press and hold the 

S

 button to confirm the hyperglycemia 

level and turn the meter off.

 Note

If the test result is greater than the pre-

set hyperglycemia level, the meter will 

display ‘Hi’.  

If the test result is greater than between 

13.4 and 33.3 mmol/L, ‘KET’ will blink 

three times on the screen.

 Note

•  If your blood glucose is measured to be more than 13.3 mmol/L, you may 

also have ketones. 

•  The CareSens S Fit BT meter is not intended to detect ketones. 

•  If you see ‘KET’ displayed on your meter, ask your healthcare professional 

when and how you should test for ketones.

 Caution

Ask your healthcare professional to help you decide what your hyperglycemia 

level is before setting your level.

 Note

If the test result is within the selected normal blood glucose 

range, the smile symbol will be displayed as shown.

14  Setting the Strip Expiration Date Indicator

Step 1 

Entering the Expiration Date Setting

Press and hold the 

 and 

 buttons at the same time for 3 seconds to 

enter the expiration date settings. After all segments flash across the 

screen, ‘EP’ will show up.

 Note

The strip expiration date is printed on the test strip vial.

Step 2 

Setting the Year

A number indicating the year will blink in the left corner of 

the screen. Press the 

 or 

 button until the correct year 

appears.

Press the 

S

 button to confirm the year and set the month.

Step 3 

Setting the Month

A number indicating the month will blink at the bottom of 

the screen. Press the 

 or 

 Press button until the correct 

month appears. Press and hold the 

S

 button for 3 seconds to 

confirm the month and turn off the meter.

15  Checking the System

You may check your meter and test strips using the CareSens S 

Glucose Control Solution (control A and/or B). The CareSens S 

Control Solution contains a known amount of glucose and is used 

to check that the meter and the test strips are working properly.

The test strip vials have CareSens S Control Solution ranges printed 

on their labels. Compare the result displayed on the meter to the 

CareSens S Control Solution range printed on the test strip vial.

Before using a new meter or a new vial of test strips, you may 

conduct a control solution test following the procedure on pages 

17–19.

 Note

•  Use only the CareSens S Control Solution (available for purchase separately).

•  Check the expiration date printed on the bottle. When you first open a 

control solution bottle, record the discard date (date opened plus three (3) 

months) in the space provided on the label.

•  Make sure your meter, test strips, and control solution are at room 

temperature before testing. Control solution tests must be done at room 

temperature 20–25 °C.

•  Before using the control solution, shake the bottle, discard the first few 

drops and wipe the tip clean.

•  Close the control solution bottle tightly and store at a temperature between 

8–30 °C.

You may do a control solution test:

•  When you want to practice the test procedure using the control solution 

instead of blood,

•  When using the meter for the first time,

•  Whenever you open a new vial of test strips,

•  If the meter or test strips do not function properly,

•  If your symptoms are inconsistent with the blood glucose test results and you 

feel that the meter or test strips are not working properly,

•  If you drop or damage the meter.

Control Solution Testing

Step 1 

Inserting Test Strip 

Insert a test strip into the meter’s test strip port, with 

the contact bars facing upwards.

Gently push the test strip into the port until the meter 

beeps. Be careful not to bend the strip while pushing 

it in.

The 

 symbol will be displayed on the screen.

Step 2 

Activating Control Solution Test Mode

You can flag the control solution test result by pressing 

and holding the 

 button for  

3 seconds. To undo the control solution flag, press and 

hold the 

 button for 3 seconds again.

Step 3 

Applying Control Solution to Test Strip 

Shake the bottle before each test. Remove the cap and 

squeeze the bottle to discard the first drop. Then wipe 

the tip with a clean tissue or cloth. Dispense a drop of 

control solution onto a clean non-absorbent surface. 

It helps to squeeze a drop onto the top of the cap as 

shown. After the 

 symbol appears on the display, 

apply the solution to the tip of the test strip until the 

meter beeps. Make sure the confirmation window fills 

completely. 

 Note

The meter may switch off if the control solution sample is not applied 

within 2 minutes of the 

 symbol appearing on the screen. If the 

meter turns off, remove the strip, reinsert, and start from step 1.

Step 4 

Waiting for the Result

The display segments will rotate clockwise and a test result will appear after the 

meter counts down from 6 to 1.

When flagged, the result is stored in the meter’s memory but it is not included 

in the averages.

01  Important Information: Read This First! -------------------------------------------- 2

02 Specifications ------------------------------------------------------------------------------ 3

03   CareSens S Fit BT Blood Glucose Monitoring System --------------------------- 4

04  Inserting or Replacing the Battery --------------------------------------------------- 5

05  Caring for Your System ------------------------------------------------------------------ 6

06  CareSens S Blood Glucose Test Strip ------------------------------------------------ 6

07  CareSens S Fit BT Blood Glucose Meter -------------------------------------------- 8

08   CareSens S Fit BT Blood Glucose Meter Display ---------------------------------- 9

09  Setting Up Your System ---------------------------------------------------------------- 10

10  Setting the Sound On/OFF ------------------------------------------------------------ 13

11  Turning on the Strip Expiration Date Indicator ---------------------------------- 13

12  Setting the Hypoglycemia (Lo) Indicator ------------------------------------------ 14

13  Setting the Hyperglycemia (Hi) Indicator ----------------------------------------- 15

14  Setting the Strip Expiration Date Indicator --------------------------------------- 16

15  Checking the System ------------------------------------------------------------------- 16

16  Using the Lancing Device -------------------------------------------------------------20

17  HI and Lo Messages ---------------------------------------------------------------------25

18  Target Blood Glucose Ranges --------------------------------------------------------26

19  Transferring Test Results --------------------------------------------------------------26

20  Meter Memory ----------------------------------------------------------------------------26

21  Setting the Alarm Function -----------------------------------------------------------28

22  Understanding Error Messages ------------------------------------------------------30

23  General Troubleshooting --------------------------------------------------------------32

24  Performance Characteristics ---------------------------------------------------------32

25  Warranty Information ------------------------------------------------------------------36

Meaning of Symbols Used: 

CE Mark
Authorised representative 

in the European 

Communmity/European 

Union

In vitro

 diagnostic medical 

device 

Caution

Biological risks  

Do not re-use

Consult instructions for use

WEEE (waste electrical and 

electronic equipment

Temperature limit

Manufacturer

Batch code

Serial number

Use-by date

•  Glucose in blood samples reacts with the chemical in the test strip to produce 

a small electrical current. The CareSens S Fit BT meter detects this electrical 

current and measures the amount of glucose in the blood sample. 

•  The CareSens S Fit BT blood glucose meter is designed to minimise code 

related errors in monitoring by using the no-coding function.

•  The CareSens S Fit BT blood glucose meter should be used only with the 

CareSens S Test Strips.

2

5

8

4

7

3

6

9

10

13

16

11

14

17

12

15

18

Summary of Contents for CareSens S Fit

Page 1: ...ting Up Your System 10 10 Setting the Sound On OFF 13 11 Turning on the Strip Expiration Date Indicator 13 12 Setting the Hypoglycemia Lo Indicator 14 13 Setting the Hyperglycemia Hi Indicator 15 14 S...

Page 2: ...screening for diabetes or for testing newborns Meaning of Symbols Used CE Mark Authorised representative in the European Communmity European Union In vitro diagnostic medical device Caution Biologica...

Page 3: ...patients should not be tested with blood glucose meters Inaccurate results may occur in patients undergoing oxygen therapy If you need assistance please contact the Spirit Healthcare 24 7 Freephone S...

Page 4: ...lood Glucose Monitoring System CareSens S Fit BT Blood Glucose Monitoring System includes the following items CareSens S Fit BT Blood Glucose Meter Instructions for Use CareSens S Fit BT Blood Glucose...

Page 5: ...ke sure the meter is turned off Push the cover in the direction of the arrow to open the battery compartment Step 2 Remove the used battery Slip your thumb finger and pulling it out with your index an...

Page 6: ...the meter components in the carrying case to prevent loss and help keep the meter clean Avoid getting any liquid or moisture in the test strip vial This can affect the test strips and cause inaccurat...

Page 7: ...est strips are used and stored according to its storage and handling methods Store test strips in a cool and dry place at a temperature between 1 30 C and 10 90 relative humidity Keep test strips away...

Page 8: ...m the meter to a computer with a cable Display Shows results and messages S Button Turns the meter on off confirms menu selections and changes information Button Turns the meter on selects or changes...

Page 9: ...e selected normal blood glucose range 11 Lo appears when the test result is lower than the selected hypoglycemia level 12 alarm appears when the time alarm has been set 13 mmol L mg dL unit for measur...

Page 10: ...s the S button when the screen display bT while OFF is blinking on the botton of the screen Then the meter will go to Step 4 Adjusting the Date and Time mode Note Follow steps 2 3 to pair your meter a...

Page 11: ...instructions on SmartLog mobile app to begin searching your meter Look for CareSens and the last 4 characters of the meter serial number on the SmartLog mobile app screen to correctly identify your m...

Page 12: ...d off after 3 seconds Then repeat Step 2 to 5 to connect your meter with smartphone Note Some smartphones especially those that are not tested or approved by i SENS may be incompatible with your meter...

Page 13: ...me for a pre set blood glucose test If the sound is set to OFF none of the sound functions will work After setting the sound press the S button to go to the next step Note The symbol is displayed only...

Page 14: ...allows you to select the desired level for the hypoglycemia indicator possible low blood sugar You will be alerted any time your test result is lower than the selected level Press the or button until...

Page 15: ...set hyperglycemia level the meter will display Hi If the test result is greater than between 13 4 and 33 3 mmol L KET will blink three times on the screen Note If your blood glucose is measured to be...

Page 16: ...nk at the bottom of the screen Press the or Press button until the correct month appears Press and hold the S button for 3 seconds to confirm the month and turn off the meter 15 Checking the System Yo...

Page 17: ...ore at a temperature between 8 30 C You may do a control solution test When you want to practice the test procedure using the control solution instead of blood When using the meter for the first time...

Page 18: ...t surface It helps to squeeze a drop onto the top of the cap as shown After the symbol appears on the display apply the solution to the tip of the test strip until the meter beeps Make sure the confir...

Page 19: ...the test result falls outside of the range Out of range results may occur in following situations Situations Do This When the control solution bottle was not shaken well When the meter test strip or t...

Page 20: ...lancing device Cap Lancet Holder Sliding Barrel Release Button Load Confirmation Window Ejection System Lancet Adjustable Tip The lancing device is for use by a single user only and should not be shar...

Page 21: ...fter use Replace lancing device tip 3 4 5 6 7 Step 4 Turn the adjustable tip until it is aligned with the load confirmation window and release button as shown 1 2 3 4 5 6 7 Step 5 The lancing device h...

Page 22: ...e sample size Preparing the Meter and Test Strip Step 7 Insert a test strip with the contact bars facing upwards into the meter s test strip port Push the strip in gently until the meter beeps Be care...

Page 23: ...allow any foreign substances such as dirt blood or water enter into the meter The meter may be damaged or may malfunction Follow the warning information provided below to prevent possible damage to t...

Page 24: ...g to a result to indicate particular situations while the strip is still in the meter When the result is displayed right after a test press the or button to select a pre meal flag a post meal flag or...

Page 25: ...3 3 mmol L and indicate severe hyperglycemia much higher than normal glucose levels If HI is displayed again upon retesting please contact your healthcare professional immediately Lo Message Lo appear...

Page 26: ...er can be transferred from the meter to a computer using SmartLog software and cable The meter screen displays PC when it is connected to the computer using the data cable For more information contact...

Page 27: ...rage values of Pre meal Post meal or fasting each representative symbol will be appeared on the screen with the average values and the number of tests peformed Step 3 Use the button to scroll back thr...

Page 28: ...et alarms alarm 1 3 The PP2 alert goes off 2 hours after setting the alert The alarms ring for 15 seconds and can be silenced by pressing the or S button or by inserting a test strip Setting the Post...

Page 29: ...is blinking on the screen On pressing the button the alarm 1 is set and On is displayed on the screen Press the button again to cancel the alarm 1 The OFF will blink on the screen Step 2 Press the bu...

Page 30: ...e appears before applying the blood or control solution sample The temperature during the test was above or below the operating range Move to an area where the temperature is within the operating rang...

Page 31: ...a problem with Bluetooth communication Contact the Spirit Healthcare 24 7 Freephone Support Line at 0800 881 5423 or go to www spirit healthcare co uk An electronic error occurred during the test Repe...

Page 32: ...w test strip The test result does not match the way you feel Repeat the test with a new test strip Check the expiration date of the test strip Perform control solution test Note If the problem is not...

Page 33: ...00 System accuracy results for glucose concentration 5 55 mmol L Within 5 Within 10 Within 15 298 420 71 0 403 420 96 0 417 420 99 3 System accuracy results for glucose concentrations between 1 5 mmol...

Page 34: ...The effect of various interfering substances was evaluated in whole blood samples The presence of the following substances within the given concentrations does not affect blood glucose measurements H...

Page 35: ...ylate 5 07 mmol L 20 Tolbutamide 3 70 mmol L 21 Tolazamide 3 21 mmol L 22 Triglycerides 37 26 mmol L 23 Uric acid 1 49 mmol L 24 Xylose 19 98 mmol L User Performance Evaluation A study evaluating gluc...

Page 36: ...at is covered by this warranty As a matter of warranty policy i SENS will not reimburse the consumer s purchase price Obtaining Warranty Service To obtain warranty service you must return the defectiv...

Reviews: